9 results match your criteria: "Kasugai CyberKnife Rehabilitation Hospital[Affiliation]"

Purpose: Lung tumor tracking during stereotactic radiotherapy with the CyberKnife can misrecognize tumor location under conditions where similar patterns exist in the search area. This study aimed to develop a technique for bone signal suppression during kV-x-ray imaging.

Methods: Paired CT images were created with or without bony structures using a 4D extended cardiac-torso phantom (XCAT phantom) in 56 cases.

View Article and Find Full Text PDF

Purpose: This study aimed to compare fiducial markers used in CyberKnife treatment in terms of metal artifact intensity observed in CT images and fiducial recognition in the CyberKnife system affected by patient body thickness and type of marker.

Methods: Five markers, ACCULOC 0.9 mm × 3 mm, Ball type Gold Anchor (GA) 0.

View Article and Find Full Text PDF

Background: There are some motion platforms for radiotherapy quality assurance. However, no platform with two drive systems that can move along three axes is available.

Purpose: The purpose of this study is to develop a dynamic motion platform with two drive systems capable of three-axis motion and to evaluate its motion performance.

View Article and Find Full Text PDF

Purpose: The present study evaluated the short-term characteristics (<3 days) of a hydrogel spacer from the time of injection during stereotactic body radiation therapy (SBRT) for prostate cancer.

Methods And Materials: Fifteen patients treated with SBRT via the CyberKnife system (36.25 Gy/5 fractions) were enrolled in this retrospective study.

View Article and Find Full Text PDF

The purpose of this study was to evaluate the effect of a hydrogel spacer on intrafractional prostate motion during CyberKnife treatment. The retrospective study enrolled 24 patients (with the hydrogel spacer = 12, without the hydrogel spacer = 12) with two fiducial markers. Regarding intrafractional prostate motion, the offset values (mm) of three axes (X-axis; superior [+] to inferior [-], Y-axis; right [+] to left [-], Z-axis; posterior [+] to anterior [-]) obtained from fiducial markers position between a digitally reconstructed radiographs images and live images in the Target Locating System were used, and extracted from generated log files.

View Article and Find Full Text PDF

Purpose: The purpose of this study was to evaluate the safety and feasibility of transarterial fiducial marker implantation for CyberKnife radiotherapy to treat locally advanced pancreatic cancer.

Materials And Methods: Fifteen pancreatic cancer patients were enrolled for transarterial marker implantation. Embolization platinum coils were implanted as a fiducial marker within 20 mm of the cancer edge, and preferably within 3 mm.

View Article and Find Full Text PDF

This study aimed to evaluate the rectal dose reduction with hydrogel spacer in 3D conformal radiotherapy (3DCRT), volumetric modulated arc therapy (VMAT), helical tomotherapy (HT), CyberKnife (CK) and proton therapy. Twenty patients who had hydrogel spacer for prostate radiotherapy were retrospectively enrolled. Computed tomography (CT) images with or without hydrogel spacer were used to evaluate rectal dose reduction.

View Article and Find Full Text PDF

Aim: The aim of the study was to evaluate computed tomography (CT) artifacts and image recognition of the CyberKnife system. Regarding fiducial markers, VISICOIL of 0.5 mm × 5.

View Article and Find Full Text PDF