Istituto di Chimica Fisica Applicata, Facolta' di Farmacia Universita' di Parma, Parma, Italy.
Homogeneous sets of data from strong acid-strong base potentiometric titrations in a mixed solvent medium (9:1 v/v methanol/water), performed by means of the glass electrode, at various constant ionic strengths and with different reference electrodes, have been analysed by statistical criteria. A standardized procedure has been established to obtain reliable potentiometric data in mixed solvents. It has been demonstrated how, with the aid of a proper linearized model, analysis of variance (ANOVA) applied to the standardization titrations can be used to test the reliability of a potentiometric chain in a medium other than pure water.