The negative interference caused by haemolysis in manual bilirubin assays contrasts with the positive interference reported for some automated methods utilizing the same basic chemistry. A comparison was therefore made of the haemolysis interference associated with several automated bilirubin methodologies: multilayer film (Kodak Ektachem- total bilirubin (TBil), direct bilirubin (DBil), conjugated bilirubin (Bc), unconjugated bilirubin (Bu),; tableted reagents (Baxter Paramax- TBil, DBil); continuous flow (Technicon SMAC - TBil). Thirty serum pools were analysed (five concentrations of bilirubin, 2-229 mumol/l; six concentrations of haemoglobin, 0.