Background: This report details our initial experience with two types of endovascular grafts- one for the treatment of infrarenal abdominal aortic aneurysms and the other for the treatment of iliac artery occlusive disease.

Methods: An abdominal aortic aneurysm was repaired in 34 patients using 3 different types of Ancure endografts (Menlo Park, California). Control patients (n = 9) had a standard aneurysm repair. Patients with chronic lower extremity ischemia (n = 7) secondary to iliac artery occlusive disease were treated with a Hemobahn endograft (W. L. Gore & Associates, Flagstaff, Arizona).

Results: Ancure graft deployment was achieved in 33 of 34 (97.1%) patients. Perioperative mortality for the Ancure and control group patients was 2.9% and 0%, respectively. Periprosthetic leaks were identified within 48 hours of deployment in 6 (18.2%) Ancure graft patients. All but 2 of the leaks resolved on serial follow-up. Additional endovascular procedures were required in 11 (32.4%) Ancure graft patients at the initial procedure or during follow-up to correct graft or arterial stenoses. Patients treated with an endovascular graft had significantly less blood loss and shorter hospital stays than control group patients. For Hemobahn patients, the technical success for graft deployment was 100%. There were no perioperative deaths. The ankle/brachial index increased from a mean of 0.52 preoperatively to 0.86 postoperatively (P = 0.004). One patient required a Wallstent in follow-up to correct a graft wrinkle. Angiography at 6 months demonstrated mild intimal hyperplasia in the stent graft in 5 of 6 patients.

Conclusions: These early results support the potential for endovascular grafts in the treatment of aneurysmal and occlusive vascular disease. Further modifications in the devices and deployment techniques are necessary to reduce the incidence of periprosthetic leaks, graft limb stenoses, and intimal hyperplasia.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0002-9610(98)00266-9DOI Listing

Publication Analysis

Top Keywords

ancure graft
12
patients
10
graft
9
aneurysmal occlusive
8
occlusive vascular
8
vascular disease
8
abdominal aortic
8
iliac artery
8
artery occlusive
8
graft deployment
8

Similar Publications

Serious Failure Modes After EVAR Are Device Specific.

J Endovasc Ther

May 2024

Division of Vascular Surgery, University of Pittsburgh, Pittsburgh, PA, USA.

Objectives: Type I and III endoleaks following endovascular aneurysm repair (EVAR) can lead to catastrophic events that require major re-interventions. We reviewed our experience with aortic endograft re-interventions for type I and III endoleaks and other serious failures among different devices.

Methods: We retrospectively reviewed patients with a prior EVAR who underwent open conversion (OC) or major endovascular intervention (MEI) (re-lining, cuff/limb extension, parallel graft) for type I/III endoleaks at our institution from 2002 to 2019.

View Article and Find Full Text PDF

Background: To evaluate the aneurysm-related complications and device issues in patients who underwent partial endograft explantation during late conversion of endovascular aneurysm repair (EVAR) to open repair.

Methods: A retrospective analysis was performed on patients who had partially explanted endografts during late conversion between 1999 and 2012. Medical records were reviewed for patient demographics, subsequent operations, and aneurysm-related complications.

View Article and Find Full Text PDF

Objective: There have been four eras in the development of endovascular aneurysm repair (EVAR): physician-made grafts, early industry devices, intermediary commercial endografts, and modern stent grafts. This study analyzes differences in outcomes between these four groups and the impact of device evolution and increased physician experience.

Methods: From 1992 to 2012, 1380 patients underwent elective EVAR.

View Article and Find Full Text PDF

Objective: With more than a decade of use of endovascular aneurysm repair (EVAR), we expect to see a rise in the number of failing endografts. We review a single-center experience with EVAR explants to identify patterns of presentation and understand operative outcomes that may alter clinical management.

Methods: A retrospective analysis of EVARs requiring late explants, >1 month after implant, was performed.

View Article and Find Full Text PDF

Background: Despite advances in endoluminal salvage for failed endografts, certain circumstances necessitate open endovascular abdominal aneurysm repair (EVAR) conversion. We review the indications for and outcomes after late EVAR explants.

Methods: Retrospective review of EVAR patients requiring delayed (>30 days) conversion from 1999 to 2009.

View Article and Find Full Text PDF