Objective: The Sometec Dynemo-3000 system allows the permanent measurement of descending aorta diameter by an echographic (A-scan) device and the blood flow velocity by a pulse Doppler velocimeter. The Dynemo-3000 then furnishes a new hemodynamic parameter, i.e., descending aortic blood flow (ABF), which is a fraction of the cardiac output (CO). We evaluate the ability of this system to measure the aortic diameter and to accurately detect ABF changes.

Design: A case study prospective trial.

Setting: A 24-bed medical intensive care unit of a 1,100-bed university hospital.

Patients: Twenty critically ill patients fully sedated, mechanically ventilated, and monitored by a pulmonary artery catheter.

Interventions: CO values determined by conventional thermodilution method (TD-CO) and ABF were recorded during the study, which included two initial baseline periods, a dobutamine infusion (5 microg/kg/min) interval of 30 mins, and a third baseline period. To assess the accuracy of A-scan, aortic diameter was measured by transesophageal echocardiography. The difference between echocardiography and A-scan was used to determine bias and precision for aortic diameter measurements. TD-CO and ABF variations were analyzed using Kruskal-Wallis and Wilcoxon tests. Association between TD-CO and ABF values was determined by calculating the linear correlation coefficient. The ability of ABF to detect a TD-CO >6.0 L/min and its variations >13% was analyzed by determination of sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values.

Measurements And Main Results: Aortic diameter measurements by A-scan and bidimensional methods were 23.0+/-2.8 mm (SD) and 24.2+/-2.7 mm, respectively. Bias and precision were 1.1 mm and 1.4 mm (95% confidence interval: -1.9 to 3.7), respectively. During the course of dobutamine infusion, we observed a significant increase of TD-CO mean value from 6.65+/-1.53 L/min to 9.30+/-2.5 L/min (p=.0008), and a parallel and significant increase in ABF mean value from 4.34+/-1.18 L/min to 5.70+/-1.63 L/min (p= .0029). Absolute TD-CO and ABF values had a correlation coefficient of 0.80. For detection of an increased TD-CO, PPV and NPV were 87% and 86%, respectively. For detection of TD-CO changes >13%, PPV and NPV were 80% and 94%, respectively.

Conclusions: The Dynemo-3000 system is able to display the real aortic diameter, which is one of the most important components of this noninvasive ultrasonic technique. When compared with TD-CO, the ABF determination provided by this ultrasonic device constitutes a reliable noninvasive tool for estimating CO and tracking its changes.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00003246-199812000-00043DOI Listing

Publication Analysis

Top Keywords

aortic diameter
20
td-co abf
20
blood flow
12
dynemo-3000 system
12
abf
9
td-co
9
cardiac output
8
aortic blood
8
sometec dynemo-3000
8
values determined
8

Similar Publications

Introduction: Abdominal aortic aneurysms (AAA) are major causes of morbidity and mortality in the elderly population. Endovascular aneurysm repair (EVAR) is associated with lower complications rates than conventional treatment; however, rigorous follow-up with contrast imaging is required to confirm aneurysmal sac exclusion. The main objective of this study was to quantify and evaluate miRNA expression response to EVAR based on serum dosages at the 6-month follow-up.

View Article and Find Full Text PDF

Abnormal characteristics of inferior vena cava and abdominal aorta among neonates with early onset septic shock.

Ital J Pediatr

January 2025

Department of Neonatology, Shanghai Children's Medical Center, School of Medicine, Shanghai Jiao tong University, Shanghai, China.

Background: The variety of shocks in neonates, if not recognized and treated immediately, is a major cause for fatality. The use of echocardiography may improve assessment and treatment, but its reference values across gestational age (GA) and birth weight (BW) are lacking. To address the information gap, this study aimed at correlating GA and BW of newborns with nonhemodynamic abnormalities, and at evaluating the usefulness of such reference values in neonates with early onset septic (EOS) -shock.

View Article and Find Full Text PDF

Objective: To explore the anatomical and clinical factors that affect the radiographic exposure time in radial artery cerebral angiography and to establish a model.

Method: A total of 210 patients who underwent radial artery cerebral angiography at this center from September 2021 to May 2022 were selected, and their anatomical and clinical factors were analyzed to evaluate the correlation between these factors and the duration of radiographic exposure. A related neural network prediction model was established.

View Article and Find Full Text PDF

Fenestrated/branched endovascular aortic repair emerges as the primary therapeutic modality for intricate aortic pathologies encompassing the paravisceral and thoracoabdominal segments, where bridging stent grafts (BSGs) play a vital role in linking the primary aortic endograft with target vessels. Bridging stent grafts can be categorized mainly into self-expanding stent grafts (SESGs) and balloon-expandable stent grafts (BESGs). Physiological factors significantly influence post-complex endovascular aortic repair BSG behaviour, impacting clinical outcomes of SESGs and BESGs in different but overlapping ways.

View Article and Find Full Text PDF

Objective: Aortic valve repair/sparing have been established as effective treatments for aortic regurgitation and/or aortic aneurysms. However, concerns remain regarding long-term durability, reproducibility, and patient selection. This study aims to asses long-term clinical and echocardiographic outcomes, with a focus on aortic regurgitation grade and left ventricular ejection fraction evolution, in adults undergoing these procedures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!