Reliability of an automatic sensing test with beat-to-beat display of the signal amplitude.

Pacing Clin Electrophysiol

Service de Cardiologie A, Centre Chirurgical Val d'Or, Saint Cloud, France.

Published: October 1998

In contrast to testing of the capture threshold, the reliability of sensing tests has been little studied. This study was performed to test the automatic sensing algorithm included in the Biotronik pacing systems. The automatic measurements made by the devices were compared with the direct manual measurements made of 271 atrial (72%) or ventricular (28%) electrograms recorded in ten patients. A high correlation (r = 0.995, P < 0.0001) was found between the two types of measurements. The use of this function should facilitate the verification of accurate sensing during the long-term follow-up of pacemaker patients, and offers a tool to analyze the variations in amplitude of intracardiac signals.