An attempt was made to evaluate the stability of an antisense oligonucleotide against nucleases present in HBL 100ras cells. To detect nanomolar concentrations of the oligonucleotide, a sensitive detection system was required. A combination of capillary electrophoresis/laser-induced fluorescence (CE-LIF) with fluorescence derivatization did not improve the sensitivity significantly and also resulted in loss of separation of the derivatized sample. On-column isotachophoresis for the preconcentration of oligonucleotide samples in DB-17 coated capillaries filled with hydroxyethyl cellulose solution could be an alternative. The isotachophoresis (ITP) step allows injection of up to 40% of the capillary volume without loss in peak resolution and peak efficiency. Using ITP-capillary polymer sieving electrophoresis (CPSE), the limit of quantitation at a signal-to-noise ratio of 10 was 73 ng/mL for a 12-mer oligonucleotide. Using these conditions, the gain in sensitivity was 125.