Purpose: To determine the minimum acceptable radiation dose for an adequate image quality in thorax a.p. radiographs of neonates using mobile X-ray equipment.

Material And Methods: The influence of additional filtration (1.0 mm Al + 0.1 mm Cu) on image quality and radiation dose was determined for the speed class 400 screen-film system (SFS) and digital luminescence radiography (DLR) by making radiographs of a test phantom. Conventional and digital thorax a.p. radiographs of a rabbit were produced using various tube current-time products. The quality of the rabbit radiographs was judged by eight radiologists applying image quality criteria according to the German guidelines and the recommendations of the European Community.

Results: The added filter resulted in a dose reduction of 39% at 66 kV. DLR gave a further dose reduction of 25% in comparison to the speed class 400 SFS while maintaining adequate image quality, i.e. the radiographs were clinically acceptable with regard to quality criteria.

Conclusion: The radiation dose resulting from thorax a.p. radiographs of neonates can be reduced by approximately 50% with the use of additional filtration and DLR.

Download full-text PDF

Source
http://dx.doi.org/10.1080/02841859809172217DOI Listing

Publication Analysis

Top Keywords

image quality
16
dose reduction
12
additional filtration
12
radiation dose
12
thorax radiographs
12
digital luminescence
8
luminescence radiography
8
adequate image
8
radiographs neonates
8
speed class
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!