The optimal field shape achieved using a multileaf collimator (MLC) often requires collimator rotation to minimize the adverse effects of the scalloped dose distribution the leaf steps produce. However, treatment machines are designed to deliver wedged fields parallel or perpendicular to the direction of the leaves. An analysis of cases from our clinic showed that for 25% of the wedged fields used to treat brain and lung tumors, the wedge direction and optimal MLC orientation differed by 20 degrees or more. The recently published omni wedge technique provides the capability of producing a wedged field with orientation independent of the orientation of the collimator. This paper presents a comparison of the three-dimensional (3D) dose distributions of the omni wedged field with distributions of wedged fields produced using both the universal and dynamic wedge techniques. All measurements were performed using film dosimetry techniques. The omni wedge generated fields closely matched the conventional wedged fields. Throughout 95% of the irradiated volume (excluding the penubra), the dose distribution of the omni wedged field ranged from +5.5 to -3.5 +/- 1.5% of that of the conventionally wedged fields. Calculation of the omni wedged field is as accurate as conventional wedged field calculation when using a 3D treatment planning systems. For two-dimensional treatment planning systems, where one must assume that the omni wedged field is identical to a conventional field, the calculated field and the delivered field differs by a small amount.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.598314DOI Listing

Publication Analysis

Top Keywords

wedged field
24
wedged fields
20
omni wedged
16
omni wedge
12
wedged
11
field
10
wedge technique
8
dose distribution
8
conventional wedged
8
treatment planning
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!