Stability of hepatitis C virus RNA during specimen handling and storage prior to NASBA amplification.

J Virol Methods

Viral Diagnostic Department, Central Laboratory of The Netherlands Red Cross Blood Transfusion Service, Amsterdam.

Published: June 1998

The influence of different anticoagulants and pre-amplification storage conditions on the stability of hepatitis C virus (HCV)-RNA, as detected by the quantitative HCV NASBA assay (NASBA-QT), was studied. The HCV-RNA load remained stable for at least 15 months when serum or plasma samples (EDTA and heparin) were directly frozen at -70 degrees C in lysis buffer. At 4 degrees C, the HCV-RNA load in serum or plasma stored with lysis buffer did not decline for at least 14 days. At 30 degrees C, however, the load declined significantly after 7 days. When clotted, whole blood was stored at 4 degrees C, the HCV-RNA load was stable for 72 h. However, when EDTA-anticoagulated whole blood was stored at 4 degrees C, the HCV-RNA load declined significantly after 48 h. In paired plasma and serum samples at baseline the HCV-RNA levels were similar. Heparin did not influence the efficiency of the HCV NASBA-QT assay. Clotted blood as well as EDTA or heparin anticoagulated blood can be used for quantifying HCV-RNA using the NASBA-QT assay. Blood samples should be stored at 4 degrees C after collection and serum or plasma separated within 24 h. Preferably, after separation, samples should be frozen in lysis buffer at -70 degrees C until NASBA-QT analysis.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0166-0934(98)00024-xDOI Listing

Publication Analysis

Top Keywords

hcv-rna load
16
serum plasma
12
lysis buffer
12
degrees hcv-rna
12
stored degrees
12
stability hepatitis
8
hepatitis virus
8
edta heparin
8
-70 degrees
8
load declined
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!