The use of improved radiochromic film for in vivo quality assurance of high dose rate brachytherapy.

Med Phys

Department of Radiation Oncology, University Hospital, SUNY at Stony Brook 11794-7028, USA.

Published: July 1998

GAFChromic film has become increasingly popular for radiation dosimetry. In this study we explore the use of GAFChromic film as an in vivo dosimeter for quality assurance (QA) of fractionated high dose rate 192Ir treatments. Accuracy of dose distribution is explored for the simple vaginal cylinder geometry for which the dose can be easily calculated for comparison. Source dwell times for several patients were optimized to deliver 500 cGy at 0.5 cm from the surface of the vaginal cylinder applicator using a commercial treatment planning system. GAFChromic film was taped to the vaginal cylinder applicator and was enclosed in a leak proof rubber sleeve prior to its insertion. Optical densities were measured along the film at 2 mm spacing, using a densitometer with filtered red light. Density corrections for transient film darkening effects were made and optical densities were converted to absorbed dose in cGy. In vivo patient dose distribution measured for different patients and different fractions were compared with the calculated values along the applicator surface. The variation between the calculated and measured dose was +/- 10%. the reproducibility of dose measurement for different fraction was within +/- 5%. This study demonstrates the potential usefulness of the film as an in vivo for brachytherapy QA.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.598311DOI Listing

Publication Analysis

Top Keywords

film vivo
12
gafchromic film
12
vaginal cylinder
12
quality assurance
8
dose
8
high dose
8
dose rate
8
dose distribution
8
cylinder applicator
8
optical densities
8

Similar Publications

Recently, there has been a great interest in the development of innovative wound dressing materials based on natural bioactives, as they can accelerate the healing process and address the issues related to traditional wound dressings. The current study focuses on developing a novel derivative of guar gum (GG) and gallic acid (GA) using a simple, free radical-mediated polymerization reaction aimed at enhancing the antioxidant properties of GG. Multiple spectroscopic investigations were performed to validate the GA-GG conjugate.

View Article and Find Full Text PDF

Urinary catheters serve as critical medical devices in clinical practice. However, the currently used urinary catheters lack efficient antibacterial and lubricating properties, often leading to discomfort with patients and even severe urinary infections. Herein, a new strategy of supramolecular assembly and disassembly of chitosan (Cs) is developed that enables efficient antibacterial lubricous and biodegradable hydrogel urinary catheters.

View Article and Find Full Text PDF

Implantation of a mesh loaded with mesenchymal stem cells (MSCs) is a common approach for the treatment of pelvic organ prolapse (POP). The mesh provides effective support to pelvic floor, enhancing muscle contraction of pelvic organs while reducing inflammation. In this study, a fully degradable mesh is designed for the treatment of POP, utilizing MSCs stimulated by a galvanic battery-powered electric field.

View Article and Find Full Text PDF

In treating type 2 diabetes, avoiding glucose reabsorption (glucotoxicity) and managing hyperglycemia are also important. A metabolic condition known as diabetes (type-2) is characterized by high blood sugar levels in comparison to normal Bilosomes (BLs) containing Dapagliflozin (Dapa) were formulated, optimized, and tested for oral therapeutic efficacy in the current investigation. Used the Box Behnken design to optimize the Dapa-BLs, formulated via a thin-film hydration technique.

View Article and Find Full Text PDF
Article Synopsis
  • Proton Minibeam Radiation Therapy has been promising in enhancing treatment efficacy compared to traditional radiation, but more research into its biological mechanisms is needed.
  • A mechanical collimation setup was developed to produce 250µm minibeams with a 1000µm spacing, with optimization using Monte Carlo simulations conducted at various proton therapy sites.
  • Results showed a peak-to-valley dose ratio (PVDR) of 10 in Dresden and 14 in Seattle, with some discrepancies between dosimetry methods that can be addressed with correction factors.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!