Purpose: Virtual colonoscopy is a new method of colon examination in which computer-aided 3D visualization of spiral CT simulates fiberoptic colonoscopy. We used a colon phantom containing various-sized spheres to determine the influence of CT acquisition parameters on lesion detectability and sizing.

Method: Spherical plastic beads with diameters of 2.5, 4, 6, 8 and 10 mm were randomly attached to the inner wall of segments of plastic tubing. Groups of three sealed tubes were scanned at 3/1, 3/2, 5/1 collimation (mm)/pitch settings in orientations perpendicular and parallel to the scanner gantry. For each acquisition, image sets were reconstructed at intervals from 0.5 to 5.0 mm. Two blinded reviewers assessed transverse cross-sections of the phantoms for bead detection, using source CT images for images for acquisitions obtained with the tubes oriented perpendicular to the gantry and using orthogonal reformatted images for scans oriented parallel to the gantry.

Results: Detection of beads of > or = 4 mm was 100% for both tube orientations and for all collimator/pitch settings and reconstruction intervals. For the 2.5 mm beads, detection decreased to 78-94% for 5 mm collimation/pitch 2 scans when the phantom sections were oriented parallel to the gantry (p = 0.01). Apparent elongation of beads in the slice direction occurred as the collimation and pitch increased. The majority of the elongation (approximately 75%) was attributable to changing the collimator from 3 to 5 mm, with the remainder of the elongation due to doubling the pitch from 1 to 2.

Conclusion: CT scanning at 5 mm collimation and up to pitch 2 is adequate for detection of high contrast lesions as small as 4 mm in this model. However, lesion size and geometry are less accurately depicted than at narrower collimation and lower pitch settings.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00004728-199807000-00028DOI Listing

Publication Analysis

Top Keywords

virtual colonoscopy
8
oriented parallel
8
collimation pitch
8
detection
5
detection colonic
4
colonic polyps
4
polyps phantom
4
phantom model
4
model implications
4
implications virtual
4

Similar Publications

Background: Adenoma detection rate and interval colon cancer rates are associated with bowel preparation quality. The US Multisociety Task Force recommends repeat colonoscopy for individuals with inadequate bowel preparation (IBP) within 1 year. However, little is known regarding the rate and associated factors of repeat colonoscopy after IBP.

View Article and Find Full Text PDF

Synthesized colonoscopy dataset from high-fidelity virtual colon with abnormal simulation.

Comput Biol Med

January 2025

State Key Laboratory of Robotics and System, Harbin Institute of Technology, Harbin, 150080, China. Electronic address:

With the advent of the deep learning-based colonoscopy system, the need for a vast amount of high-quality colonoscopy image datasets for training is crucial. However, the generalization ability of deep learning models is challenged by the limited availability of colonoscopy images due to regulatory restrictions and privacy concerns. In this paper, we propose a method for rendering high-fidelity 3D colon models and synthesizing diversified colonoscopy images with abnormalities such as polyps, bleeding, and ulcers, which can be used to train deep learning models.

View Article and Find Full Text PDF

Accurate size measurement of colorectal polyps is critical for clinical decision making and patient management. This systematic review aimed to evaluate the current techniques used for colonic polyp measurement to improve the reliability of size estimations in routine practice.A comprehensive literature search was conducted across PubMed, EMBASE, and MEDLINE to identify studies relevant to size measurement techniques published between 1980 and March 2024.

View Article and Find Full Text PDF

Background: The aim of this Delphi study was to reach consensus on a new clinical decision tool to help identify or exclude Crohn's disease (CD) in patients with perianal fistula(s) (PAF).

Methods: A panel of international experts in the field of proctology/Inflammatory Bowel Disease (IBD) were invited to participate. In the first round (electronic survey), participants were asked to anonymously provide their opinion probing 1) the relevance and use of clinical characteristics suggestive of underlying CD, 2) the use of faecal calprotectin (FCP) for screening for CD and 3) on the diagnostic work-up for CD in PAF patients with raised clinical suspicion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!