The purpose of the QSL-Finland study was to assess the state-of-the-art trueness and precision of serum total-calcium and glucose measurements in Finnish clinical laboratories. For this purpose, 21 hospitals and clinical institutes were selected. They measured six single donation sera, the total-calcium (t-calcium) and glucose content of which had been determined by ion chromatography and isotope dilution-gas chromatography-mass spectrometry (ID-GC-MS) reference methods. The results were interpreted in light of specifications for imprecision, bias and total error of routine methods that have been proposed in the past. The data revealed that the performance of t-calcium and glucose methods is generally acceptable in Finnish clinical laboratories. This study did not lead to a separation of laboratories according to accreditation. In consequence, it seems that accreditation, in its present form, cannot substitute dedicated quality assurance practices.

Download full-text PDF

Source
http://dx.doi.org/10.1080/00365519850186625DOI Listing

Publication Analysis

Top Keywords

state-of-the-art trueness
8
trueness precision
8
precision serum
8
serum total-calcium
8
total-calcium glucose
8
glucose measurements
8
measurements finnish
8
qsl-finland study
8
finnish clinical
8
clinical laboratories
8

Similar Publications

Introduction: The aim of this study was to perform a comprehensive verification of a 6-part differential haematology analyser Siemens Advia 2120i (Erlangen, Germany), prior to its routine implementation.

Materials And Methods: Our verification protocol included: precision (within- and between-run), estimated bias (%) as measure of trueness, which was calculated from observed and manufacturers' declared value, analytical measuring interval (AMI), carryover, confirmation of a limit of blank (LoB), determination of a limit of detection (LoD) and limit of quantitation (LoQ). The K ethylenediaminetetraacetic acid (EDTA) patients' leftover samples were used for verification of analyser Advia 2021i.

View Article and Find Full Text PDF

Background: DYMIND DH76 (DYMIND BIOTECH, China) is a new automated hematology system designed to provide CBC count, including a 5-part WBC differential count, and its analytical performance should be assessed before adoption for clinical use.

Methods: The analyzer was evaluated according to the International Council for Standardization in Haematology guideline. The purposes of this study were to assess its analytical performance in comparison to SYSMEX XN 1000 hematology analyzer currently used in our laboratory, as well as to compare the automated and manual WBC differential.

View Article and Find Full Text PDF

: Quantitative proteomics using mass spectrometry is performed label-free or label-based approaches. Labeling strategies rely on the incorporation of stable heavy isotopes by metabolic, enzymatic, or chemical routes. Isobaric labeling uses chemical labels of identical masses but of different fragmentation behaviors to allow the relative quantitative comparison of peptide/protein abundances between biological samples.

View Article and Find Full Text PDF

Development, validation and application of an ICP-MS/MS method to quantify minerals and (ultra-)trace elements in human serum.

J Trace Elem Med Biol

September 2018

Institute of Nutritional Science, University of Potsdam, Arthur-Scheunert-Allee 114-116, 14558 Nuthetal, Germany; NutriAct - Competence Cluster Nutrition Research, Berlin, Potsdam, Germany; TraceAge - DFG Research Unit on Interactions of Essential Trace Elements in Healthy and Diseased Elderly, Potsdam-Berlin-Jena, Germany. Electronic address:

Multi-element determination in human samples is very challenging. Especially in human intervention studies sample volumes are often limited to a few microliters and due to the high number of samples a high-throughput is indispensable. Here, we present a state-of-the-art ICP-MS/MS-based method for the analysis of essential (trace) elements, namely Mg, Ca, Fe, Cu, Zn, Mo, Se and I, as well as food-relevant toxic elements such as As and Cd.

View Article and Find Full Text PDF

Background: To provide its participants with an external quality assessment system (EQAS) that can be used to check trueness, the Dutch EQAS organizer, Organization for Quality Assessment of Laboratory Diagnostics (SKML), has innovated its general chemistry scheme over the last decade by introducing fresh frozen commutable samples whose values were assigned by Joint Committee for Traceability in Laboratory Medicine (JCTLM)-listed reference laboratories using reference methods where possible. Here we present some important innovations in our feedback reports that allow participants to judge whether their trueness and imprecision meet predefined analytical performance specifications.

Methods: Sigma metrics are used to calculate performance indicators named 'sigma values'.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!