Modular designs of hip prostheses have become popular recently. Along with complications inherent in all hip arthroplasty systems, modular systems have the additional potential for dissociation of components. Five male patients underwent total hip arthroplasties, in which all of the acetabular components were Harris-Galante II porous acetabular cups. Many years after the operation, the polyethylene liners were dissociated without any previous trauma or dislocation of the femoral heads, these dissociations and dislodgements were managed with open reduction. This complication can be predicted from clinical symptoms and signs. Roentgenograms must be taken and carefully compared to previous roentgenograms. We postulated two causes for the dissociation. First, the polyethylene liner was not fixed securely within the acetabular metal shell at the time of operation. Second, the locking mechanism of the acetabular metal shell was not strong enough to firmly hold the polyethylene liner within the acetabular metal shell. It does warrant that certain precautions must be taken when implanting modular components. The locking mechanism of the harris-Galante II porous acetabular component is mechanically weak and fails easily, therefore its design must be improved in an attempt to prevent postoperative dissociation of the polyethylene liner.

Download full-text PDF

Source
http://dx.doi.org/10.3349/ymj.1998.39.3.277DOI Listing

Publication Analysis

Top Keywords

polyethylene liner
16
acetabular metal
16
metal shell
16
dissociation polyethylene
12
total hip
8
harris-galante porous
8
porous acetabular
8
locking mechanism
8
acetabular
7
polyethylene
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!