Use of a simulator with CT option in radiotherapy of macular degeneration.

Int J Radiat Oncol Biol Phys

Department of Radiation Oncology, Academic Hospital, Free University of Brussel, Belgium.

Published: June 1998

Purpose: To assess the accuracy of a conventional simulation procedure in radiotherapy of age-related macular degeneration.

Methods And Materials: A computed tomographic (CT) extension attached to the treatment simulator was used to acquire CT images immediately after conventional simulation in 18 patients referred for treatment of age-related macular degeneration. Analysis was performed on 16 one-sided treatment cases for whom images were obtained. Error was estimated by the displacement between the observed treatment isocenter and the intended isocenter based on reconstructed eye geometry.

Results: Based on single slice measurements, the mean error amplitude was 2.3 mm (range 0.2-5.6). Based on three-dimensional eye globe reconstruction, the mean error amplitude was 2.8 mm (range 0.8-5.3). An incidental finding previously unreported was the lower image quality at the center of the simulator-CT image acquisition field.

Conclusions: Small but significant errors from conventional simulation were noted. The integrated simulation-CT procedure may help correct the errors to improve the accuracy of simulation setup. The lower image quality at the center of image acquisition field requires adaptation of the simulation-CT procedure.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0360-3016(98)00070-4DOI Listing

Publication Analysis

Top Keywords

conventional simulation
12
macular degeneration
8
age-related macular
8
error amplitude
8
amplitude range
8
lower image
8
image quality
8
quality center
8
image acquisition
8
simulation-ct procedure
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!