Long-term results of vena cava filters: experiences with the LGM and the Titanium Greenfield devices.

Cardiovasc Intervent Radiol

Institut für Roentgendiagnostik, Würzburg, Germany.

Published: September 1998

Purpose: Vena cava filter (VCF) application is the method of choice to prevent recurrent pulmonary embolism in patients with deep venous thrombosis. Because of the reported complications after VCF placement we summarize our long-term follow-up results with the LGM and Titanium Greenfield (TG) devices.

Methods: Eighty-seven LGM VCF and 17 TG VCF were placed in 104 patients (average age 64 years). The follow-up examinations were performed by color-coded duplex sonography, plain radiographs, cavography, and computed tomography (CT). The maximum observation time was 81 months.

Results: Filter migration occurred in 11% (8/76) of the LGM VCF and 15% (2/13) of the TG VCF. Vena cava thrombosis was seen in 17% (13/76) of the patients with an LGM VCF and in 31% (4/13) of those with a TG VCF. The patency rate was 95% (72/76) for the LGM VCF and 92% (12/13) for the TG VCF. Pulmonary embolism was noted in 3 patients after LGM VCF insertion and in no patient after TG VCF insertion.

Conclusion: A VCF should only be inserted in a patient after pulmonary embolism and when there is strict proof of the indication.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s002709900249DOI Listing

Publication Analysis

Top Keywords

lgm vcf
20
vcf
13
vena cava
12
pulmonary embolism
12
lgm titanium
8
titanium greenfield
8
embolism patients
8
patients lgm
8
lgm
7
long-term vena
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!