We conducted a multicenter evaluation of the analytical and clinical features of the automated Bayer Immuno 1 CA 15-3 assay and compared assay performance to two manual tests. Results of the 10-day imprecision study of the Bayer Immuno 1 assay pooled across four evaluation sites and three lots of reagent produced total CV < or = 4%. Lot-to-lot reproducibility for 26 different lots of reagents and calibrators manufactured over a 2-year period was demonstrated (CV, 1.1%). Results for the Bayer Immuno 1 assay correlated well with the Biomira TRUQUANT BR 27.29 and Centocor CA 15-3 RIAs (r > or = 0.94). The upper limit of the reference interval for the Bayer Immuno 1 assay was 35.9 kilounits/L (35.9 units/mL); values were similar for all methods. Longitudinal monitoring of healthy women yielded assay values with an average CV of 11% and 21% for the Bayer Immuno 1 and Biomira assays, respectively. The Bayer Immuno 1 assay demonstrated the analytical features, intermethod correlation, and long-term performance characteristics that are essential for longitudinal monitoring of breast cancer patients.