We prospectively studied 223 patients (288 lesions) who underwent elective or bail out implantation of 309 NIR stents (Scimed, Boston Scientific Corporation, Galway, Ireland). Most lesions (68.4%) had unfavorable characteristics (type B2 or C). Primary success in stent deployment was achieved in 305 (98.6%). There was no Q-wave myocardial infarction. Emergency coronary artery bypass grafting (CABG) was required in 1 patient and 1 death occurred. Subacute thrombosis rate was 0.4%. Reference diameter was 2.65+/-0.67 mm. Minimum luminal diameter (MLD) increased from 0.62+/-0.45 to 2.69+/-0.57 mm and diameter stenosis decreased from 78.3+/-13.4% to 12.7+/-5.9%. Clinical follow-up was performed in the first 135 patients for 5.3+/-1.6 months and repeat angiography was undertaken in 35 (16%) with recurrence of symptoms at 4.6+/-1.3 months. Clinical restenosis rate was 9.6%. We conclude that the NIR coronary stent exhibits favorable performance characteristics and appears to be safe and efficacious in the treatment of coronary lesions even in the presence of high-risk characteristics.

Download full-text PDF

Source
http://dx.doi.org/10.1002/(sici)1097-0304(199802)43:2<153::aid-ccd9>3.0.co;2-9DOI Listing

Publication Analysis

Top Keywords

nir coronary
8
coronary stent
8
preliminary experience
4
experience nir
4
coronary
4
stent prospectively
4
prospectively studied
4
studied 223
4
223 patients
4
patients 288
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!