This report documents clinical and hemodynamic benefits of balloon mitral valvuloplasty (BMV) using a bifoil balloon catheter from a single center in 415 consecutive cases of rheumatic mitral stenosis (MS). The procedure was successful in 396 (95.2%) patients, with an increase in mitral valve area (MVA) from 0.82+/-0.35 cm2 to 2.21+/-0.24 cm2 (P < 0.001). There were 2 (0.48%) in-hospital deaths, and 6 (1.44%) patients developed acute mitral regurgitation. The procedural and fluroscopy time was reduced significantly from 52+/-11 and 38+/-6 min to 33+/-7 and 19+/-5 min, respectively, after modifications of technique in our last 326 cases. The bifoil balloon catheter technique is safe and effective, and provides excellent hemodynamic benefits which are sustained at long-term follow-up. This technique should be considered as an addition to the existing armamentarium of interventional cardiologists performing mitral balloon valvuloplasty.

Download full-text PDF

Source
http://dx.doi.org/10.1002/(sici)1097-0304(199801)43:1<43::aid-ccd12>3.0.co;2-uDOI Listing