Dose distributions that result from treating a patient with orthovoltage beams are best determined with a treatment planning system that uses the Monte Carlo method, and such systems are not readily available. In the present work, the Monte Carlo method was used to develop a computer code for determining absorbed dose distributions in orthovoltage radiation therapy. The code was used in planning treatment of a patient with a neuroendocrine carcinoma of the maxillary sinus. Two lateral high-energy photon beams supplemented by an anterior orthovoltage photon beam were utilized in the treatment plan. For the clinical case and radiation beams considered, a reasonably uniform dose distribution (+/- 10%) is achieved within the target volume, while the dose to the lens of each eye is 4-8% of the prescribed dose. Therefore, an orthovoltage photon beam, when properly filtered and optimally combined with megavoltage beams, can be effective in the treatment of cancers below the skin, providing that accurate treatment planning is carried out to establish with accuracy and precision the doses to critical structures.