Comparison of two different pulse oximeters in monitoring preterm infants.

Crit Care Med

Department of Child Life and Health, University of Edinburgh, and the Neonatal Unit, Simpson Memorial Maternity Pavilion, UK.

Published: December 1997

Objective: The aim of the study was to test the reliability and variation in the readings of two widely used pulse oximeters in preterm infants.

Design: Two different pulse oximeters and a transcutaneous PO2 monitor were used to record the data continuously on a cotside computer database.

Patients: Sixteen preterm infants were studied in the Neonatal Unit, Simpson Memorial Maternity Pavilion, Edinburgh, UK.

Interventions: None.

Measurements And Main Results: Approximately one fifth of the time, the pulse oximeter readings could be established as artifactual. Study of the remaining four fifths of the data showed that, on average, the Nellcor pulse oximeter recorded saturation percentages 2.2% higher than the Ohmeda oximeter.

Conclusions: We recommend that all neonatal units adopt a policy of using different saturation alarm limits for these two instruments. We further recommend that other pulse oximeters be tested by a methodology similar to the one we present in this paper, before their use in monitoring oxygenation in preterm infants.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00003246-199712000-00025DOI Listing

Publication Analysis

Top Keywords

pulse oximeters
16
preterm infants
12
pulse oximeter
8
pulse
5
comparison pulse
4
oximeters
4
oximeters monitoring
4
preterm
4
monitoring preterm
4
infants objective
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!