We developed criteria for implantation and programming of permanent endocardial pacemakers in patients with a nonthoracotomy ICD system. These criteria were prospectively used in 10 patients who recieved an ICD prior to (n = 5) or following (n = 5) implantation of a dual chamber (n = 6) or ventricular (n = 4) pacemaker with a unipolar (n = 4) or bipolar (n = 6) lead configuration. All patients were tested for interactions or malfunctions. Undersensing of ventricular fibrillation by the atrial sense amplifier and inadequate atrial pacing occurred in one patient with a unipolar dual chamber system programmed to AAIR but didn't impair ICD sensing. Transient or permanent loss of capture or sensing of the pacemaker was not observed after ICD shocks with the output programmed to double pulse width and voltage of stimulation threshold and the sensitivity to 50% of the detected R wave. One episode of transient reprogramming occurred without clinical consequences. One unipolar ventricular pacemaker lead had to be exchanged against a bipolar lead because of oversensing of the pacing artifact by the ICD. There was no failure of an ICD to detect ventricular arrhythmias due to inadequate pacemaker activity. During a follow-up period of 21 +/- 11 months, a total of 78 ventricular arrhythmias were effectively treated in six patients. Thus, a combined use of transvenous ICD and pacemaker is possible despite the close vicinity of pacing and defibrillations leads. Optimized programming different to the common settings is required. As interactions occurred only in unipolar pacemaker leads bipolar systems should be used in these patients.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.1997.tb05435.xDOI Listing

Publication Analysis

Top Keywords

combined transvenous
8
dual chamber
8
ventricular pacemaker
8
bipolar lead
8
ventricular arrhythmias
8
icd
7
pacemaker
6
patients
5
ventricular
5
transvenous implantation
4

Similar Publications

Background: Lymphatic leaks are associated with significant mortality and morbidity. Intranodal lymphangiography (ILAG) involves the direct injection of ethiodised lipid into the hilum of lymph nodes. It is diagnostic procedure that can have therapeutic effects secondary to a local sclerosant effect.

View Article and Find Full Text PDF

Orbital arteriovenous fistulas (AVFs) are rare vascular malformations that can cause severe ocular complications. This review evaluates the effectiveness of treatment strategies, focusing on post-treatment recovery and recurrence. A systematic review was conducted using PubMed and Scopus with no date restrictions.

View Article and Find Full Text PDF

Unlabelled: New cardiac implantable electronic devices (CIEDs), such as leadless pacemakers and subcutaneous implantable cardioverter defibrillators (S-ICDs), are being used in patients with adult congenital heart disease. The selection of CIEDs often requires careful consideration due to technical challenges posed by a unique heart structure. A 27-year-old man following a surgical tetralogy of Fallot (TOF) repair developed non-sustained ventricular tachycardia, sick sinus syndrome, and complete atrioventricular block.

View Article and Find Full Text PDF
Article Synopsis
  • The study explores optimal reimplantation strategies after the removal of infected cardiovascular implantable electronic devices (CIEDs), highlighting the potential advantages of leadless pacemakers (LPs) due to their lower infection risk.
  • A literature review included 22 relevant studies, covering 657 patients who had lead removal followed by LP implantation, with 44.9% receiving LPs concurrently during the procedure and minimal complications reported.
  • The findings suggest that both concurrent and delayed LP implantation is safe, with low rates of reinfection and complications, positioning LPs as a viable option for patients needing temporary pacing before further treatments.
View Article and Find Full Text PDF

Inappropriate Therapy and Shock Rates Between the Subcutaneous and Transvenous Implantable Cardiac Defibrillator: A Secondary Analysis of the PRAETORIAN Trial.

Circ Arrhythm Electrophysiol

December 2024

Department of Cardiology, Amsterdam UMC Location University of Amsterdam, Heart Center, Amsterdam Cardiovascular Sciences Heart failure and Arrhythmias, the Netherlands (L.R.A.O.N., S.P., L.V.A.B., T.F.B., A.-F.B.E.Q., W.v.d.S., L.S., J.A.d.V., J.G.P.T., N.R.B., J.R.d.G., K.M.K., A.d.W., A.A.M.W., R.E.K.).

Article Synopsis
  • Inappropriate therapy (IAT) is a significant issue associated with implantable cardiac defibrillator (ICD) therapy, particularly highlighted in early subcutaneous ICD (S-ICD) studies which showed high rates of inappropriate shocks (IAS).
  • The PRAETORIAN trial, an international study with 849 patients, found no major differences in IAT and IAS rates between S-ICD and transvenous ICD (TV-ICD) groups, as both groups had similar cumulative incidences.
  • Key predictors for IAT varied between the two groups, with TV-ICD patients experiencing IAT mainly from supraventricular tachycardias, while S-ICD patients faced issues from cardiac oversensing
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!