Unlabelled: This study evaluates the performance of the newly developed high-resolution whole-body PET scanner ECAT EXACT HR+.

Methods: The scanner consists of four rings of 72 bismuth germanate block detectors each, covering an axial field of view of 15.5 cm with a patient port of 56.2 cm. A single block detector is divided into an 8 x 8 matrix, giving a total of 32 rings with 576 detectors each. The dimensions of a single detector element are 4.39 x 4.05 x 30 mm3. The scanner is equipped with extendable tungsten septa for two-dimensional two-dimensional measurements, as well as with three 68Ge line sources for transmission scans and daily quality control. The spatial resolution, scatter fraction, count rate, sensitivity, uniformity and accuracy of the implemented correction algorithms were evaluated after the National Electrical Manufacturers Association protocol using the standard acquisition parameters.

Results: The transaxial resolution in the two-dimensional mode is 4.3 mm (4.4 mm) in the center and increases to 4.7 mm (4.8 mm) tangential and to 8.3 mm (8.0 mm) radial at a distance of r = 20 cm from the center. The axial slice width measured in the two-dimensional mode varies between 4.2 and 6.6 mm FWHM over the transaxial field of view. In the three-dimensional mode the average axial resolution varies between 4.1 mm FWHM in the center and 7.8 mm at r = 20 cm. The scatter fraction is 17.1% (32.5%) for a lower energy discriminator level of 350 keV. The maximum true event count rate of 263 (345) kcps was measured at an activity concentration of 142 (26.9) kBq/ml. The total system sensitivity for true events is 5.7 (27.7) cps/Bq/ml. From the uniformity measurements, we obtained a volume variance of 3.9% (5.0%) and a system variance of 1.6% (1.7%). The implemented three-dimensional scatter correction algorithm reveals very favorable properties, whereas the three-dimensional attenuation correction yields slightly inaccurate results in low- and high-density regions.

Conclusion: The ECAT EXACT HR+ has an excellent, nearly isotropic spatial resolution, which is advantageous for brain and small animal studies. While the relatively low slice sensitivity may hamper the capability for performing fast dynamic two-dimensional studies, the scanner offers a sufficient sensitivity and count rate capacity for fully three-dimensional whole-body imaging.

Download full-text PDF

Source

Publication Analysis

Top Keywords

count rate
12
whole-body pet
8
pet scanner
8
national electrical
8
electrical manufacturers
8
manufacturers association
8
ecat exact
8
field view
8
spatial resolution
8
scatter fraction
8

Similar Publications

To study the risk of incident dementia after a non-traumatic intracranial hemorrhage in a diverse US population, and evaluate if this risk is different for the subtypes of intracranial hemorrhage. We performed a retrospective cohort study using both inpatient and outpatient claims data on Medicare beneficiaries between January 1, 2008 and December 31, 2018. The exposure was a new diagnosis of non-traumatic intracranial hemorrhage, defined as a composite of intracerebral hemorrhage (ICH), subarachnoid hemorrhage (SAH), and subdural hemorrhage (SDH).

View Article and Find Full Text PDF

Introduction: In high-altitude cities located above 2,500 m, hospitals face a concerning mortality rate of over 50% among intensive care unit (ICU) patients with acute respiratory distress syndrome (ARDS). This elevated mortality rate is largely due to the absence of altitude-specific medical protocols that consider the unique physiological adaptations of high-altitude residents to hypoxic conditions. This study addresses this critical gap by analyzing demographic, clinical, sex-specific, and preclinical data from ICUs in Bogotá, Colombia (2,650 m) and El Alto, Bolivia (4,150 m).

View Article and Find Full Text PDF

Objective: The C-reactive protein/Lymphocyte Ratio (CLR) is a novel biomarker whose role in the development of chronic kidney disease (CKD) is not well understood. This study aimed to investigate the correlation between CLR and the prevalence of CKD.

Methods: This cross-sectional study included participants from the US National Health and Nutrition Examination Survey conducted between 1999 and 2010.

View Article and Find Full Text PDF

Objectives: To assess the safety, reliability and acceptability of the modified incremental step test (MIST) supervised remotely via videoconferencing in adults with chronic respiratory disease.

Methods: Adults with chronic respiratory disease undertaking pulmonary rehabilitation were invited to undertake the MIST under two testing conditions: in-person supervision and remote supervision via video-conferencing. Test order was randomised.

View Article and Find Full Text PDF

Frozen versus fresh embryo transfer in women with low prognosis for in vitro fertilisation treatment: pragmatic, multicentre, randomised controlled trial.

BMJ

January 2025

State Key Laboratory of Reproductive Medicine and Offspring Health, Centre for Reproductive Medicine, Institute of Women, Children and Reproductive Health, Shandong University, Jinan, China

Objective: To test the hypothesis that a freeze-all strategy would increase the chance of live birth compared with fresh embryo transfer in women with low prognosis for in vitro fertilisation (IVF) treatment.

Design: Pragmatic, multicentre, randomised controlled trial.

Setting: Nine academic fertility centres in China.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!