Objective: The objective of the study was to compare the accuracy of the TDxFLM test (Abbott Laboratories) with the fetal lung maturity cascade (shake, foam stability index, lecithin/sphingomyelin tests) and to determine whether the TDxFLM test could increase the efficiency and reduce the cost without decreasing the reliability of a cascade.
Study Design: A prospective, single-blinded study was conducted. Uncontaminated amniotic fluid obtained by transabdominal amniocentesis for fetal lung maturity assessment was evaluated with use of the fetal lung maturity cascade and the TDxFLM test. At study completion the results of the TDxFLM test were compared with those of the maturity cascade with regard to hyaline membrane disease, which was defined by strict clinical and radiographic parameters. A power analysis was performed requiring a sample size of 100 infants delivered within 72 hours of amniocentesis with use of the 95% confidence interval.
Results: A total of 115 cases had a full maturity cascade performed, of which 40 (35%) had a positive shake or foam stability index and 75 cases required progression to a lecithin/sphingomyelin ratio because of negative results. The TDxFLM test result was > or = 70 mg/gm in 42 (37%) of these 115. One hundred eight newborns were delivered within 72 hours of the amniocentesis; 65% (71) of these were between 30 and 37 weeks of estimated gestational age. There were 7 cases of hyaline membrane disease in the 108 newborns. Of these 108, 87 had a mature original cascade versus 85 mature tests with use of a proposed TDxFLM test-lecithin/sphingomyelin ratio cascade with one case of respiratory distress syndrome and hyaline membrane disease. The sensitivity, specificity, and positive and negative predictive values for the original cascade were 86%, 84%, 27%, and 99%, respectively; for the proposed TDxFLM test-lecithin/sphingomyelin ratio cascade the values were 86%, 83%, 26%, and 99%, respectively. The TDxFLM test-lecithin/sphingomyelin ratio cascade would have resulted in a cost reduction of 24% with no significant delay in turnaround time.
Conclusion: The TDxFLM test appears to be a reliable and accurate assessment of fetal lung maturity. Furthermore, by replacing the shake and foam stability index portion of the cascade with the TDxFLM test, a cost savings of 24% would occur without a decrease in safety. These results also reveal that it could enhance patient care and be cost efficient for institutions not currently doing fetal pulmonary maturity testing to undertake use of the TDxFLM test and to only send out specimens for a lecithin/sphingomyelin ratio that have an initial immature TDxFLM test result (< 70 mg/gm). Likewise, institutions currently only performing a lecithin/sphingomyelin ratio may consider a TDxFLM test-lecithin sphingomyelin ratio cascade. Although direct costs would increase, they would be counterbalanced by a significant reduction in laboratory technician time.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0002-9378(97)70278-3 | DOI Listing |
Am J Obstet Gynecol
December 2012
Department of Maternal-Fetal Medicine, Institut Clínic de Ginecologia, Obstetrícia i Neonatologia, Hospital Clinic, Universitat de Barcelona, Barcelona, Spain.
Objective: The objective of the study was to evaluate the performance of automatic quantitative ultrasound analysis (AQUA) texture extractor to predict fetal lung maturity tests in amniotic fluid.
Study Design: Singleton pregnancies (24.0-41.
J Matern Fetal Neonatal Med
August 2012
Department of Obstetrics and Gynecology, Hospital of the University of Pennsylvania, Philadelphia, PA 19104, USA.
Objective: To determine if infants delivered after immature or indeterminate TDx-FLM II testing and a mature reflex test are at increased risk for neonatal respiratory complications.
Methods: The primary analysis compared neonatal respiratory morbidity (RDS or TTN) in 34-39-week fetuses delivered after either (i) mature TDx-FLM II testing, or (ii) indeterminate or immature TDx-FLM II and a positive reflex test (PG or L/S ratio).
Results: Fifty patients delivered after mature TDx-FLM II, and 30 after immature or indeterminate TDx-FLM II with an L/S ≥ 2.
Rev Bras Ginecol Obstet
March 2010
Purpose: To compare the lamellar body number density (LBND) count in amniotic fluid using the fluorescent polarization (FP) test as a diagnostic parameter for the assessment of fetal pulmonary maturity.
Method: This was an analytical, controlled cross-sectional study conducted on 60 pregnant women from March 2002 to December 2007. Amniotic fluid specimens were obtained by amniocentesis or at the time of caesarean section, and submitted to the LBND and FP tests (TDxFLM, Abbott Laboratories), the latter considered to be a reference test, and compared in terms of the presence or absence of respiratory distress syndrome (RDS).
Fetal Diagn Ther
December 2009
Department of Maternal-Fetal Medicine, Hospital Clínic, Institut d'Investigacions Biomèdiques Augusto Pi i Sunyer, University of Barcelona, and Centro de Investigación Biomédica en Red de Enfermedades Raras, Barcelona, Spain.
Aim: To construct a predictive model for respiratory distress syndrome (RDS) from gestational age (GA) at delivery and TDx-FLM II value.
Methods: Pregnant women who underwent an amniocentesis in which TDx-FLM II was determined were included in the study. A model for the occurrence of RDS was constructed by means of a logistic regression procedure from TDx-FLM II values and GA at delivery.
Am J Obstet Gynecol
November 2005
Department of Obstetrics and Gynecology, Washington University School of Medicine, St Louis, MO, USA.
Objective: The purpose of this study was to determine the mean weekly increment in amniotic fluid TDx-FLM II ratio during the latter part of pregnancy.
Study Design: All women who underwent > 1 amniocentesis for the determination of fetal lung maturity between 1998 and 2004 were identified retrospectively. Clinical information and TDx-FLM II ratios were collected from the participant's chart and analyzed.