We investigated the effect of collimator selection on image quality in regional cerebral blood flow (rCBF) studies of the brain performed with 99Tc(m)-HMPAO. A triple-headed SPET system (GE/CGR Neurocam) was used, together with three sets of parallel-hole collimators - general-purpose (GP), high-resolution (HR) and ultra-high-resolution (UHR). Two image quality parameters were used to describe the image quality, namely, noise and resolution. Noise was measured in experimental and Monte-Carlo simulated SPET studies of a cylinder phantom of uniform activity as the pixel root mean square error (RMS) and as the coefficient of variation (CV) of quantitative rCBF values. Resolution was measured as full-width at half-maximum in experimental SPET studies of a line-source. Plots of noise versus resolution for the different collimators were obtained by varying the cut-off frequency of the Hanning filter applied in the reconstruction of transaxial slices. From these noise-resolution plots, we were able to determine which collimator gave the best resolution for a specific noise level. A lowest reasonable noise level may be established by comparison with the inter-observer CV of the quantification method.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00006231-199707000-00010DOI Listing

Publication Analysis

Top Keywords

image quality
12
rcbf studies
8
triple-headed spet
8
noise-resolution plots
8
spet studies
8
noise level
8
noise
5
selection collimator
4
collimator rcbf
4
studies
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!