Determination of total cholesterol in serum by liquid chromatography-isotope dilution mass spectrometry.

Clin Chem

Institute for Clinical Chemistry and Pathobiochemistry, Medical Faculty, University of Technology Aachen, Germany.

Published: October 1997

We have developed a liquid chromatography-isotope dilution mass spectrometry procedure to quantify total cholesterol in serum. A particle-beam interface was used for coupling the liquid chromatograph and the mass spectrometer. After electron impact ionization the ions m/z = 386 and m/z = 389 were used for selective ion monitoring of cholesterol and the internal standard [25,26,27-(13)C]cholesterol. The sample preparation steps required for serum materials are alkaline hydrolysis and an extraction of the cholesterol into the cyclohexane phase. Imprecision for the determination of cholesterol in control materials is typically <1.0%. The deviation from the certified reference values was <0.75% for all control materials tested. A method comparison of the results obtained by this method with those obtained by gas chromatography-isotope dilution mass spectrometry for n = 28 pooled human sera derived from samples analyzed in our routine laboratory did not show differences >2.5%.

Download full-text PDF

Source

Publication Analysis

Top Keywords

total cholesterol
8
cholesterol serum
8
liquid chromatography-isotope
8
chromatography-isotope dilution
8
dilution mass
8
mass spectrometry
8
cholesterol
5
determination total
4
serum liquid
4
spectrometry developed
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!