Purpose: The authors present the results of a prospective study of 125 consecutive cases of phacoemulsification, performed by a single surgeon, to assess the benefit of computerised videokeratoscopy for the improvement of the predictive calculation of the power of intraocular lenses (IOL).

Methods: Preoperative keratometry was determined by means of 3 concurrent techniques, including automated kerato-refractometry (AKR 2000, Nidek) and videokeratoscopy by TMS-1 (Tomey) and CAS-1 (EyeSys). The results of the IOL calculation by these 3 methods was compared postoperatively to the power of the actual IOL yielding emmetropia as calculated by the optical method based on the postop, refraction at the various time-points.

Results: Simulated values obtained with videokeratoscopy significantly reduced the refractive error at 1 postop, week (p = 0.001), one month (p = 0.002) and 3 months (p = 0.002) for the CAS1 and at 3 months (p = 0.002) for the TMS1, as compared to values obtained with conventional automated keratometry. The refractive error was significantly reduced with the CAS1 at one month (p = 0.001) and 3 months (p = 0.019) as compared to TMS1.

Conclusion: The determination of the anterior curvature of the entrance pupil by videokeratoscopy significantly improves the predictive calculation of the IOL power.

Download full-text PDF

Source

Publication Analysis

Top Keywords

predictive calculation
12
calculation power
8
power intraocular
8
refractive error
8
months 0002
8
[computerized video-keratoscopic
4
video-keratoscopic analysis
4
analysis cataract
4
cataract surgery
4
surgery phacoemulsification
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!