Purpose: To evaluate different ultrasound modalities during implantation and follow-up of endovascular grafts for abdominal aortic aneurysm (AAA) exclusion.
Methods: Between February 1995 and May 1996, 18 patients (14 men; aged 49 to 80 years, mean 67) were treated with endovascular intervention for infrarenal AAA. Seventeen patients received Mialhe Stentor bifurcated grafts, while one patient was treated with a straight graft for pseudoaneurysm. During and after the implantation, 3.25- and 5-MHz annular array ultrasound probes were used for transabdominal visualization of the endograft. Intravascular ultrasound was applied in combination with angiography for postoperative control.
Results: Intraprocedurally, transabdominal two-dimensional (2D) ultrasound successfully monitored guidewire passage from the groin into the main part of the bifurcated endograft for implantation of the second limb. All implantation procedures were technically successful, but four endoleaks were identified intraprocedurally by 2D ultrasound and angiography. One healed spontaneously, two were treated with endovascular techniques at 1 and 4 months, and the last leak was scheduled for repair when the patient died of probable myocardial infarction at 2 months. During follow-up, 2D ultrasound successfully visualized all the endografts; no endoleaks were found in up to 18 months of surveillance.
Conclusions: Transabdominal ultrasound imaging could be valuable in bifurcated endograft deployment both for guiding guidewire insertion and for controlling wire position before the second graft limb is connected to the main graft. Provided that satisfactory visualization of the entire endograft can be obtained, ultrasound examination may possibly replace arteriography and computed tomographic scanning as a follow-up investigation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1583/1074-6218(1997)004<0272:UIDEAA>2.0.CO;2 | DOI Listing |
J Endovasc Ther
January 2025
Vascular Unit, Department of Surgery, Mater Dei Hospital, Msida, Malta.
Purpose: The use of surgeon-modified fenestrated endograft to treat a bleeding complication in the common iliac artery.
Technique: An Endurant limb graft was modified on back table in theater after planning the fenestration using a semi-automated centerline. The Endurant stent was planned to land flush at the aortic bifurcation.
J Endovasc Ther
December 2024
Department of Vascular and Endovascular Surgery, Klinik Ottakring, Wiener Gesundheitsverbund, Wien, Austria.
Objective: This study offers a retrospective assessment of a single-center experience using cerebrospinal fluid catheters to reduce the risk of perioperative spinal cord injury in patients undergoing single-staged complex endovascular juxtarenal or thoracoabdominal aortic aneurysm repair.
Results: A total of 97 patients were included. On average, 70.
Int J Vasc Med
November 2024
Second Department of Surgery, Division of Vascular Surgery, "G. Gennimatas" General Hospital of Thessaloniki, School of Health Sciences, Faculty of Medicine, Aristotle University of Thessaloniki, Thessaloniki, Greece.
J Vasc Surg Cases Innov Tech
February 2025
Department of Vascular and Endovascular Surgery, Semmelweis Aortic Center, Heart and Vascular Center, Semmelweis University, Budapest, Hungary.
Ann Vasc Surg
November 2024
Department of Public Health, Vascular Surgery Unit, University Federico II of Naples, Naples, Italy.
Background: The AFX unibody endograft offers advantages in treating abdominal aortic aneurysms (AAAs) with narrow aortic bifurcations due to its unique design, where the flow divider sits at the native aortic bifurcation. However, its limited length options may limit complete proximal neck utilization. As per device instruction for use, the implant should be completed with an aortic cuff.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!