A feasibility study for mailed film dosimetry has been performed. The global reproducibility of the method is better than 2%. It is shown that the normalized sensitometric curve does not depend on photon beam quality in the range from Co-60 gamma-rays to 18 MV x-rays, although the dose per optical density decreases when the energy increases. The fading of the latent image before film processing is only 3% per month and the normalized sensitometric curve is not modified after a period of 51 days between irradiation and processing. Sets of films were mailed to three different institutes for irradiation and returned for processing and evaluation after more than two months in order to verify that mailing of irradiated and unprocessed films does not produce unwanted artefacts. Finally the feasibility of external audits with mailed film dosimetry is illustrated by comparison of beam profiles measured with films and ionization chambers in a polystyrene phantom.