Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
The Mycobacterium tuberculosis chaperonin 10 protein and fragments corresponding to sequences 59-99, 51-99 and 26-99 were synthesised by the solid-phase methodology using a double coupling protocol and without the aid of capping agents. After the final acid cleavage using the low TFMSA-high HF protocol the polypeptides were purified by either the ion exchange chromatography/RP-HPLC combination or the isoelectric separation carried out in solution and followed by semi-preparative RP-HPLC. Comparison of the results obtained through the two approaches indicated that in general the isoelectricfocusing/HPLC combination was superior both in terms of recovery of final material and its purity. The advantages found were as follows: (i) Unlike ion exchange chromatography, no tailoring of the separation conditions is required, (ii) Several consecutive focusings can be carried out in progressively narrower pH gradients. This increases the separation resolution without the need of changing other separation parameters, (iii) Very little manipulation is needed, and each focusing requires 3-5h. (iv) Full compatibility with non-ionic denaturants such as 8 M urea. This increases solubility so that using the ROTOFOR instrument described here 50-100 mg crude polypeptide can be processed daily. Thus the isoelectric focusing technique carried out in solution is a valid and inexpensive alternative to ion exchange chromatography.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1399-3011.1997.tb01131.x | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!