US extended-field-of-view imaging technology.

Radiology

Ultrasound Group, Siemens Medical Systems, Issaquah, WA 98029-7002, USA.

Published: June 1997

Purpose: To develop an ultrasound (US) extended-field-of-view scanning technique that combines the convenience of a real-time scanner with the spatial advantages of a static B-mode scanner and provides a panoramic image in real time without position sensors or cumbersome articulated arms.

Materials And Methods: An image-registration-based position-sensing technique was used to track probe motion and reconstruct a large composite image during real-time scanning. The probe motion (translation and rotation) was estimated by combining multiple local motion vectors. This computationally intensive process required a special programmable image processor.

Results: Large, resolution-preserved composite images up to 60 cm long were obtained. Measurement accuracy as determined with phantom experiments was better than 5%. The method could be applied to any probe or image format.

Conclusion: In addition to providing a panoramic image to expand diagnostic capabilities, extended-field-of-view US provides a more easily interpretable image and is an effective cross-specialty communication tool.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiology.203.3.9169720DOI Listing

Publication Analysis

Top Keywords

panoramic image
8
probe motion
8
image
6
extended-field-of-view imaging
4
imaging technology
4
technology purpose
4
purpose develop
4
develop ultrasound
4
ultrasound extended-field-of-view
4
extended-field-of-view scanning
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!