This study was designed to test the sensitivity and specificity of a combination of 3 phospholipid-dependent assays performed with various reagents, for the detection of lupus anticoagulant (LA). Plasmas containing an LA (n = 56) or displaying various confounding pathologies [58 intrinsic pathway factor deficiencies, 9 factor VIII inhibitors, 28 plasmas from patients treated with an oral anticoagulant (OAC)] were selected. In a first step, the efficiency of each assay and reagent was assessed using the Receiving Operating Characteristic (ROC) method. Optimal cut-offs providing both sensitivity and specificity > or = 80% were determined. The APTT assay and most of the phospholipid neutralization assays failed to discriminate factor VIII inhibitors from LA. In a second step, using the optimal cut-offs determined above, the results of all the possible combinations of the 3 assays performed with 4 different reagents were analyzed. Thirteen combinations of reagents allowed > or = 80% of plasmas of each category (LA, factor deficiency or OAC) to be correctly classified (3/3 positive test results in LA-containing plasmas and 0/3 positive results in LA-negative samples).