[Vertebral morphometry with DXA/MXA equipment--an equipment comparison. Dual x-ray absorptiometry/morphometric x-ray absorptiometry].

Rofo

Abteilung für Radiologische Diagnostik und Nuklearmedizin, Universitätsklinikum Benjamin Franklin, Freie Universität, Berlin.

Published: February 1997

Purpose: The accuracy and reproducibility of morphometric measurements (Morphometric X-ray Absorptiometry = MXA) of vertebrae were determined. The significant difference of the change in height of vertebral bodies in follow-up studies was computed for MXA methods and digitised spinal radiographs as well.

Material And Methods: The measurements were carried out on two new Dual X-ray Absorptiometry (DXA) devices (device A = Expert, Lunar Corp., device B = QDR 2000 Plus, Hologic Inc.). The data were obtained by using the European Spine Phantom (ESP) and lumbar spine specimens.

Results: The accuracy of vertebral morphometry performed on radiographs is 2.0%, on the device A 2.3%, and on the device B 4.9%. Measurements taken with the ESP showed a reproducibility of 1.0 to 3.0%, whereas measurements of fractured vertebrae resulted in 5.1 to 6.0%.

Conclusion: The results of the morphometric measurements demonstrate that a reliable fracture analysis in phantoms and specimens is possible. Further in-vivo studies are necessary.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2007-1015396DOI Listing

Publication Analysis

Top Keywords

dual x-ray
8
morphometric measurements
8
x-ray absorptiometry
8
measurements
5
[vertebral morphometry
4
morphometry dxa/mxa
4
dxa/mxa equipment--an
4
equipment--an equipment
4
equipment comparison
4
comparison dual
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!