The Captia Syphilis IgG enzyme immunoassay (EIA) was evaluated for use in conjunction with the rapid plasma reagin test (RPR) as a method to test for syphilis. A total of 1,288 serum specimens were tested by the routine laboratory protocol of the RPR followed by microhemagluttination assay for Treponema pallidum (MHA-TP) testing of RPR-reactive sera as well as the EIA-RPR protocol in which the automated EIA followed by a manual RPR test for EIA-positive specimens is used. When using the routine protocol, 131 specimens were initially reactive by the RPR, and 113 of these were reactive by MHA-TP. When using the EIA-RPR protocol, 170 specimens were initially positive by EIA, and of these, 112 were RPR reactive, indicating active disease. When compared to the routine protocol, the EIA-RPR protocol had sensitivity, specificity, and positive and negative predictive values of 96.5, 99.7, 97.3, and 99.7%, respectively. After resolution of discrepancies by additional testing, the adjusted sensitivity, specificity, and positive and negative predictive values were 100, 99.8, 98.3, and 100%, respectively. This evaluation demonstrates that when used in conjunction with the RPR, the Captia Syphilis EIA is a reliable method by which to test for syphilis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC232718PMC
http://dx.doi.org/10.1128/jcm.35.5.1141-1143.1997DOI Listing

Publication Analysis

Top Keywords

captia syphilis
12
test syphilis
12
eia-rpr protocol
12
syphilis igg
8
conjunction rapid
8
rapid plasma
8
plasma reagin
8
reagin test
8
method test
8
routine protocol
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!