[Influence of visible interferences on biochemical assays realised on Au 5231, AU 5223 (Olympus) and CL 7200 (Shimadzu)].

Ann Biol Clin (Paris)

Laboratoire de biochimie automatisée, CHU de Nancy-Brabois, Vandaeuvre-lès-Nancy, France.

Published: April 1997

The impact of the major interferents (hemolysis, bilirubin, turbidity), on the quality of biochemical tests, was evaluated on multiparametric analysers (CL 7200 Shimadzu, Japan/Ciba-Corning, France; AU 5231 and AU 5223 Olympus, Japan/bioMérieux, France), according to the SFBC instructions. Interferences were detected in 33 cases upon 165 tests realized, that is to say 20% of the performed analysis. Turbidity was the most frequent cause of interference (7.8%), followed by hemolysis (8.5%) and bilirubin (3.6%). The use of a sample blank, a bireagent, the change of reagent, the change of the secondary wavelength or the modification of the measurement times, allowed us to reduce more than 80% of the interferences. Only three interferences remained: hemolysis upon the measurement of TGO and potassium, and bilirubin upon the measurement of creatinine. For these parameters, a suitable note using the Olympus factors (semi quantitative expression of the importance of the three interferents) is reported on the answer sheet.

Download full-text PDF

Source

Publication Analysis

Top Keywords

5231 5223
8
5223 olympus
8
[influence visible
4
interferences
4
visible interferences
4
interferences biochemical
4
biochemical assays
4
assays realised
4
realised 5231
4
olympus 7200
4

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!