A sensitive and selective method for the determination of cefuroxime in bronchoalveolar lavage (BAL) fluid using high-performance liquid chromatography (HPLC) with UV detection at 280 nm after solid-phase extraction with C18 cartridges was developed. A Waters symmetry C18 column was used and the mobile phase was acetonitrile-0.05 M ammonium phosphate buffer (pH 3.2) (15:85, v/v). The method enabled the determination of cefuroxime at concentrations below 100 ng/ml, with a linear calibration curve at concentrations of 5-100 ng/ml for 400 microliters of BAL. The intra- and inter-assay coefficient of variations for 10, 40 and 80 ng/ml were between 5.3 and 8.9%. Analytical recoveries were between 92.7 and 106.2%. The detection limit was 1 ng/ml at a signal-to-noise ratio of 3:1 using 400 microliters of BAL. The method was successfully used for the analysis of BAL fluid from patients after oral administration of 500 mg cefuroxime axetil twice daily.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0378-4347(96)00329-5 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!