AI Article Synopsis

  • The study tested the effect of initial phase polarity on the Defibrillation Threshold (DFT) in two types of implanted defibrillators (ICDs) in patients with ventricular tachycardia or fibrillation.
  • Patients receiving the Medtronic PCD 7219C showed significantly lower DFT with one polarity (RV+) compared to the other (RV-), while the PCD 7219D group showed a similar trend that wasn't statistically significant.
  • Overall, the results suggest that adjusting the initial phase polarity of the biphasic shock waveform during ICD implantation can influence the DFT, highlighting its importance in optimizing defibrillator performance.

Article Abstract

The effect of initial phase polarity on the DFT of two pectorally implanted biphasic ICDs was tested in a randomized, prospective manner at the time of implantation. Twenty-two consecutive patients with VT-VF who received either the Medtronic PCD 7219C Jewel device (10 patients) or PCD 7219D Jewel device (12 patients) were studied. DFT testing was performed in a standard step-down manner. Both initial phase polarities--initial defibrillation current flowing from active can/SVC coil +/- subcutaneous patch) to the RV coil (RV coil to active can/SVC (RV+)--were tested in random order. The mean DFT achieved with RV+ compared with RV-was lower for the 7219C patient group (6.6 +/- 3.1 vs 10.8 +/- 5.5 J; P = 0.007). A similar trend was observed for the 7219D group, though the difference did not reach statistical significance (12.0 +/ 4.0 vs 16.3 +/- 7.3 J; P = 0.07). Seven of the 10 patients in the 7219C group had a lower DFT with RV+, while the initial phase polarity made no difference in 3. In the 7219D group, 7 patients had a lower DFT using RV+, 2 patients had a lower DFT using RV-, and the initial phase polarity made no difference in 3. In conclusion, this study demonstrates that changing the polarity of the initial phase of a biphasic shock waveform can have a significant impact on the DFT achieved at the time of ICD implantation.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.1997.tb06178.xDOI Listing

Publication Analysis

Top Keywords

initial phase
24
phase polarity
12
lower dft
12
polarity initial
8
phase biphasic
8
biphasic shock
8
shock waveform
8
pectorally implanted
8
jewel device
8
device patients
8