Effect of different experimental parameters on the potentiometric evaluation of blood electrolytes using K+ as a test cation.

Analyst

CECUL-Centro de Electroquímica e Cinética de Universidade de Lisboa, Portugal.

Published: December 1996

AI Article Synopsis

  • The study examines how different methods of preparing albumin solutions impact the measurement of potassium ion (K+) concentrations in solutions containing albumin with KCl.
  • It highlights that the sequence of measurements and the preparation processes (dilution vs. concentration) influence the relationship between electrical potential (E) and albumin concentration, showing variations based on whether the solutions are hypertonic or isotonic.
  • The results suggest that careful consideration of preparation methods is essential, as they can meaningfully affect the liquid-liquid junction potential and overall findings in potentiometric evaluations.

Article Abstract

Potentiometric evaluation of ionic concentrations in physiological media has been reported to be significantly affected by the albumin containing matrix. Previous studies have attempted to clarify the origin of different patterns of variation of E versus albumin concentration at a constant 0.01 mol dm-3 KCl, depending on the experimental methodologies. This paper reports on measurements of K+ in albumin (BSA) containing KCl solutions, which have been pursued following different methodologies concerning the sequence of measurements, i.e., starting either from the most concentrated solution (100 g dm-3) (A) or from the most dilute solution (20 g dm-3) (B), and solution preparation, (a) dilution of an initial 100 g dm-3 albumin solution by successive addition of 0.01 mol dm-3 KCl, (b) concentration of an initial 20 g dm-3 albumin solution by addition of solid substance, (c) independent preparation of each albumin solution (20, 40, 60, 80 and 100 g dm-3) or (d) preparation of each albumin solution from direct dilution of a 100 g dm-3 stock solution with 0.01 mol dm-3 KCl. Convenient calculations were made and showed a significant contribution of albumin to the liquid-liquid junction potential. The variation of potential with albumin concentration is smaller for hypertonic than for isotonic bridge solutions, both for the dilution series (A,a and A,d) and for the concentration series (B,c and B,d). When the method for increasing the concentration is performed by addition of a solid substance to an initially diluted albumin (B,b), the slope of the E versus albumin concentration plot is larger for hypertonic than for isotonic reference electrolytes. This latter finding is in agreement with Payne's results when ultrafiltration was used to concentrate a protein solution. These observations are a clear indication that the method of preparation of concentrated protein solutions may significantly affect the results.

Download full-text PDF

Source
http://dx.doi.org/10.1039/an9962101907DOI Listing

Publication Analysis

Top Keywords

100 dm-3
16
albumin solution
16
albumin concentration
12
001 mol
12
mol dm-3
12
dm-3 kcl
12
albumin
11
dm-3
9
solution
9
potentiometric evaluation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!