Dose reduction in double contrast barium enema by use of low fluoroscopic current.

Br J Radiol

Department of Diagnostic Imaging, St James's Hospital, Dublin, Ireland.

Published: September 1996

Under standard conditions double contrast barium enema (DCBE) was carried out on a selected patient cohort using either standard (1.5 mA) or low (0.5 mA) fluoroscopic current. A statistically significant (p < or = 0.01) reduction in total dose was achieved by use of low screening current. This reduction was of the order of 40%. Fluoroscopic image quality was analysed both objectively and subjectively, and was found not to be significantly degraded when the lower current was used. On blinded evaluation all examinations were found to be of diagnostic quality. The use of low fluoroscopic current results in a reduction in patient dose, without compromising fluoroscopic image quality or diagnostic standard of the examination. Low fluoroscopic current technique has been adopted as the standard in our department.

Download full-text PDF

Source
http://dx.doi.org/10.1259/0007-1285-69-825-852DOI Listing

Publication Analysis

Top Keywords

low fluoroscopic
16
fluoroscopic current
16
double contrast
8
contrast barium
8
barium enema
8
current reduction
8
fluoroscopic image
8
image quality
8
fluoroscopic
6
current
6

Similar Publications

Background: Traditional freehand puncture relies on non-real-time computed tomography (CT) images, which significantly affects the accuracy of puncturing targets in the lower lung lobes with respiratory motion. This study aims to assess the safety and feasibility of a teleoperated robotic system and low-dose CT for the accurate real-time puncture of targets in the lungs of live pigs during breathing under fluoroscopic guidance.

Methods: Two puncture methods were analyzed: freehand and robot-assisted.

View Article and Find Full Text PDF

: Spinal flexibility radiographs are important in adolescent idiopathic scoliosis (AIS) for clinical decision-making. In this study, we introduce a new method, the 'quantitatively controlled standing fulcrum side-bending' test (CSFS test). This is a feasibility study; we aimed to quantify the applied force and track the temporospatial changes in the spine specifically by measuring the continuous change in the Cobb angle (in degrees) during lateral bending.

View Article and Find Full Text PDF

Clinical feasibility study of transcatheter edge-to-edge mitral valve repair in dogs with the canine V-Clamp device.

Front Vet Sci

December 2024

Department of Clinical Sciences and James L. Voss Veterinary Teaching Hospital, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, CO, United States.

Objective: To determine procedural feasibility, safety, and short-term efficacy in dogs with severe degenerative mitral regurgitation (MR) undergoing transcatheter edge-to-edge repair (TEER) with a canine-specific device.

Design: Prospective, single-arm (uncontrolled), single-institution clinical feasibility study.

Animals: Fifty client-owned dogs with severe degenerative MR operated over a 28-month period.

View Article and Find Full Text PDF

A 78-year-old female with a remote history of L3-4 decompression and fusion presented with several months of low back and radicular leg pain. MRI revealed moderate L2-L3 spinal canal stenosis, ligamentum flavum infolding, moderate bilateral foraminal stenosis, and a grade I retrolisthesis. A right sided L2-L3 TFESI was performed using multiplanar fluoroscopic imaging with a subpedicular supraneural approach.

View Article and Find Full Text PDF
Article Synopsis
  • - The study investigates the average fluoroscopic time and radiation exposure during basivertebral nerve radiofrequency ablations (BVNRFA), a new treatment for chronic low back pain that utilizes fluoroscopy for precise probe placement.
  • - A total of 55 patients were analyzed; the findings revealed an average fluoroscopic time of 152.5 seconds, with radiation exposure quantified as a cumulative dose of 70.3 mGy.
  • - The results highlight that the average fluoroscopic time required for BVNRFA is approximately 2.5 minutes, contributing valuable data on radiation safety during this procedure.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!