Background And Aims Of The Study: Wear of Björk-Shiley Delrin (BSD) heart valve discs is known to have occurred in some patients, possibly contributing to increased regurgitation. This paper specifically addresses surface and edge wear that have been observed on some discs of explanted BSD valves after implant durations up to 22.4 years.

Methods: The wear patterns have been documented using either photographic or scanning electron microscopic methods for 42 out of 73 explanted BSD valve discs. The remainder of the 73 discs were not available for analysis.

Results: One form of surface wear found on 18 out of 42 of the Delrin discs was the concentric wear of mild abrasive origin along the surface near the disc edge due to contact with the inlet and outlet struts. In five instances, surface anomalies were observed, primarily in the areas of high velocity blood flow. This paper also describes two Delrin discs with non-concentric edge wear patterns: (a) one which appears to be due to fatigue micro-chipping and abrasive wear of the disc of a 20 year BSD explant, which had fibrous tissue ingrowth, causing abnormal rotation of the disc during valve closure, and (b) a second one which is thought to have been caused by a cutting action of the knife-like stub of one inlet strut leg which had separated. Cross-sectional analyses of two explanted BSD discs, with full indent grooves on the inflow side, indicated that the Delrin material was primarily compressed under these wear grooves, rather than removed by abrasion. Hardness profiles indicated that the Delrin 150 microns below the surface was harder and would tend to prevent further deformation. A simple model describing the compound impact (impact with sliding) phenomenon is introduced to explain abrasive wear found on some explanted BSD discs.

Conclusion: Based on the studies here and reports in the literature, the BSD heart valve appears to present a design which provides many years of service and, when wear occurs, it occurs in a manner that provides easily recognized clinical symptoms, which allow time for diagnosis and treatment.

Download full-text PDF

Source

Publication Analysis

Top Keywords

explanted bsd
16
edge wear
12
heart valve
12
valve discs
12
wear
11
surface edge
8
wear björk-shiley
8
björk-shiley delrin
8
discs
8
bsd heart
8

Similar Publications

Donation After Circulatory Death Heart Transplantation: A Narrative Review.

J Cardiothorac Vasc Anesth

September 2024

Department of Anesthesia and Critical Care Medicine, University of Chicago, Chicago, IL, United States.

Heart transplantation is the definitive treatment for refractory, end-stage heart failure. The number of patients awaiting transplantation far exceeds available organs. In an effort to expand the donor pool, donation after circulatory death (DCD) heart transplantation has garnered renewed interest.

View Article and Find Full Text PDF

Male versus female inflammatory response after brain death model followed by ex vivo lung perfusion.

Biol Sex Differ

January 2024

Laboratorio de Cirurgia Cardiovascular e Fisiopatologia da Circulação (LIM-11), HC-FMUSP, Instituto Do Coração (InCor), Faculdade de Medicina da Universidade de São Paulo, Av. Dr. Arnaldo, 455 2º Andar, Sala 2146, São Paulo, 01246-903, Brazil.

Background: Ex vivo lung perfusion (EVLP) is a useful tool for assessing lung grafts quality before transplantation. Studies indicate that donor sex is as an important factor for transplant outcome, as females present higher inflammatory response to brain death (BD) than males. Here, we investigated sex differences in the lungs of rats subjected to BD followed by EVLP.

View Article and Find Full Text PDF

Introduction: Preeclampsia (PE) is a serious hypertensive pregnancy disorder and a leading cause of maternal and perinatal morbidity and mortality. Despite the prevalence and complications, there are no approved therapeutics to relieve PE symptoms. Inflammation, oxidative stress, and angiogenic imbalance have been shown to contribute to the PE pathophysiology, though there is a lack of understanding in how best to target these pathways in PE.

View Article and Find Full Text PDF

Commensal bacteria signal through TLR5 and AhR to improve barrier integrity and prevent allergic responses to food.

Cell Rep

October 2023

Department of Pathology, The University of Chicago, Chicago, IL 60637, USA; Pritzker School of Molecular Engineering, The University of Chicago, Chicago, IL 60637, USA; Committee on Immunology, The University of Chicago, Chicago, IL 60637, USA. Electronic address:

The increasing prevalence of food allergies has been linked to reduced commensal microbial diversity. In this article, we describe two features of allergy-protective Clostridia that contribute to their beneficial effects. Some Clostridial taxa bear flagella (a ligand for TLR5) and produce indole (a ligand for the aryl hydrocarbon receptor [AhR]).

View Article and Find Full Text PDF

Orthotopic heart transplantation with persistent left superior vena cava draining into coronary sinus: the anterior rerouting technique.

Gen Thorac Cardiovasc Surg

April 2021

Department of Surgery, Section of Cardiac and Thoracic Surgery, Center for Aortic Diseases, The University of Chicago Medicine, 5841 S Maryland Avenue, MC5040, Chicago, IL, 60637, USA.

Background: We present a surgical technique for a reconstruction of a persistent left superior vena cava draining into the coronary sinus with orthotopic heart transplantation.

Methods: Between January 2013 and July 2019, a total of 224 patients underwent orthotopic heart transplantation in our institution. Among them, three patients (3/224, 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!