The volume of current implantable cardioverter defibrillators (ICD) is not convenient for pectoral implantation. One way to reduce the size of the pulse generator is to find a more effective defibrillation pulse waveform generated from smaller volume capacitors. In a prospective randomized crossover study we compared the step-down defibrillation threshold (DFT) of a standard biphasic waveform (STD), delivered by two 250-microF capacitors connected in series with an 80% tilt, to an experimental biphasic waveform delivered by a single 450-microF capacitor with a 60% tilt. The experimental waveform delivered the same energy with a lower peak voltage and a longer duration (LVLD). Intraoperatively, in 25 patients receiving endocardial (n = 12) or endocardial-subcutaneous array (n = 13) defibrillation leads, the DFT was determined for both waveforms. Energy requirements did not differ at DFT for the STD and LVLD waveforms with the low impedance (32 +/- 4 omega) endocardial-subcutaneous array defibrillation lead system (6.4 +/- 4.4 J and 5.9 +/- 4.2 J, respectively) or increased slightly (P = 0.06) with the higher impedance (42 +/- 4 omega) endocardial lead system (10.4 +/- 4.6 J and 12.7 +/- 5.7 J, respectively). However, the voltage needed at DFT was one-third lower with the LVLD waveform than with the STD waveform for both lead systems (256 +/- 85 V vs 154 +/- 51 V and 348 +/- 76 V vs 232 +/- 54 V, respectively). Thus, a single capacitor with a large capacitance can generate a defibrillation pulse with a substantial lower peak voltage requirement without significantly increasing the energy requirements. The volume reduction in using a single capacitor can decrease ICD device size.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8159.1996.tb03312.x | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!