A method to determine the activity of the cytochrome P-450 1A1 enzyme, by measuring 7-ethoxyresorufin-O-deethylase (EROD) activity using high-performance liquid chromatography (HPLC) with fluorescence or with visible absorbance detection of resorufin, is described. The lowest quantifiable activity (0.2 pmol/mg min) is obtained by incubation of 0.3 mg of human duodenal microsomal proteins using HPLC fluorescence detection. Using HPLC with visible absorbance detection, sensivity was ten times lower. However, the equipment for this last method is available in most laboratories. The use of both HPLC assays allows determination of the low EROD activity level in samples of small size, such as two or three human duodenal biopsies obtained by routine endoscopy. These methods will be a useful tool to study the role of drug intestinal metabolism by cytochrome P-450 1A1.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0378-4347(95)00543-9DOI Listing

Publication Analysis

Top Keywords

cytochrome p-450
12
p-450 1a1
12
visible absorbance
12
absorbance detection
12
activity high-performance
8
high-performance liquid
8
liquid chromatography
8
fluorescence visible
8
erod activity
8
hplc fluorescence
8

Similar Publications