Fracture of a left atrial catheter made of polyethylene and its subsequent embolization and entrapment in a Bjork-Shiley aortic valvular prosthesis is reported in a 62-year-old woman. The entrapped fragment of the catheter was retrieved by a transarterial approach using a homemade modified snare-loop device. Measures to prevent such a complication should be directed towards avoidance of force when removing such catheters. In addition, their length should be determined prior to insertion and again after removal, in order to detect their possible fracture and loss at once.

Download full-text PDF

Source
http://dx.doi.org/10.1378/chest.72.1.98DOI Listing

Publication Analysis

Top Keywords

fragment catheter
8
bjork-shiley aortic
8
aortic valvular
8
valvular prosthesis
8
transarterial approach
8
removal fragment
4
catheter entrapped
4
entrapped bjork-shiley
4
prosthesis transarterial
4
approach modified
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!