This paper describes CaSO4:Dy and LiF:Mg, Cu, P which were used for ambient environmental monitoring before the nuclear power plant operation in Guangdong Daya Bay, China, in 1991. Since LiF:Mg, Cu, P was first used as an environmental dosimeter in this laboratory, the intercomparison of both thermoluminescent dosimeters, including laboratory irradiation and environmental exposure in Beijing reference spots, was conducted in cooperation with National Institute of Metrology and Laboratory of Industrial Hygiene, measured values of both thermoluminescent dosimeters were in agreement with the error being less than +or- 2% for the laboratory irradiation. The results of measurement by both thermoluminescent dosimeters were quite in agreement with environmental reference exposure rates measured by a pressurized ionization chamber. The largest error of CaSO4:Dy environmental monitoring results in Daya Bay also showed that the differences of measurement results between two thermoluminescent dosimeters were not significant. The experiment results indicated that LiF:Mg, Cu, P was a good environmental dosimeter.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/00004032-199603000-00007 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!