A common problem with low-energy electron beams (< 15 MeV) is their low surface dose when the incident electrons are monodirectional. This makes it difficult to deliver a uniform dose to tumor with any precision, limiting the clinical usefulness of such beams. A practical method is presented for greatly increasing the tissue depth enclosed by the 95% isodose region, while delivering the entire dose in a single uninterrupted treatment. Beam modification is achieved by placing a wire mesh of high atomic number (Z) on the treatment surface throughout the treatment. Electron beams of energies 6, 9, and 13 MeV are modified to produce a near uniform dose from the surface to the original depth of maximum, approximately doubling the depth enclosed by the 95% isodose. These beams have a step-function-like depth dose and can be arranged to deliver a constant dose to tumor located at varying depths while simultaneously sparing deeper tissues which are also located at varying depths. A single mesh design was found to be suitable for all energies.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.597432DOI Listing

Publication Analysis

Top Keywords

electron beams
12
depth dose
8
wire mesh
8
uniform dose
8
dose tumor
8
depth enclosed
8
enclosed 95%
8
95% isodose
8
located varying
8
varying depths
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!