AI Article Synopsis

  • Previous research indicates that the polarity of epicardial patches affects the success of defibrillation, but its impact on biphasic shocks with endocardial electrodes is unclear.
  • The study involved 20 patients using an Endotak lead system to determine the defibrillation threshold for biphasic pulses, assessing different polarities of the electrodes during the process.
  • Results showed that 60% of patients had a lower defibrillation threshold when the proximal coil was negative, suggesting that changing the polarity of biphasic shocks can enhance defibrillation efficacy and potentially reduce the need for more invasive procedures.

Article Abstract

Background: Previous studies have shown that the polarity of epicardial patches significantly affects the defibrillation efficacy of monophasic shocks. However, whether this improvement can be extended to different pulsing methods and lead systems, such as biphasic shocks using endocardial defibrillating electrodes, is unknown.

Methods: Twenty consecutive patients undergoing testing and permanent implant using an Endotak lead system with a biphasic device were included in the study. In each patient the defibrillation threshold was determined delivering biphasic pulses with the distal coil as the cathode and the proximal coil as the anode during the positive phase and with the polarity reversed. The initial electrode polarity tested was chosen randomly. The defibrillation threshold was defined as the lowest pulse amplitude that effectively terminated ventricular fibrillation induced with 60-Hz alternating current. For each biphasic pulse peak voltage, pulse duration, resistance, and stored energy were recorded.

Results: Of the 20 patients, 12 (60%) had lower defibrillation threshold when the proximal coil was negative, whereas only 2 patients had a lower defibrillation threshold when the distal coil was negative. In four patients a subcutaneous patch would have been required if only the biphasic pulse with the distal coil as negative had been tested. The mean stored defibrillation threshold energy was lower with the configuration using the proximal coil as cathode (16.3 +/- 8.8 J vs 21.5 +/- 11 J; P < 0.01).

Conclusion: Change in the initial polarity of biphasic shocks may influence defibrillation efficacy and should, therefore, be assessed in each patient to achieve a more satisfactory safety margin and minimize the use of more invasive lead configurations.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.1995.tb03837.xDOI Listing

Publication Analysis

Top Keywords

defibrillation threshold
24
distal coil
12
proximal coil
12
coil negative
12
initial polarity
8
defibrillation
8
biphasic pulses
8
defibrillation efficacy
8
biphasic shocks
8
coil cathode
8

Similar Publications

Background: Implantable cardioverter-defibrillators (ICDs) are essential for reducing sudden cardiac death in patients at risk of ventricular arrhythmias. The choice of ICD lead - single-coil or dual-coil - can influence device performance and patient outcomes. This meta-analysis evaluates the comparative efficacy and safety of single-coil versus dual-coil ICD leads to inform clinical decision-making.

View Article and Find Full Text PDF

Background: Patients with a left ventricular ejection fraction ≤ 35% are at increased risk of sudden cardiac death (SCD) within the first months after a myocardial infarction (MI). The wearable cardioverter defibrillator (WCD) is an established, safe and effective solution which can protect patients from SCD during the first months after an MI, when the risk of SCD is at its peak. This study aimed to evaluate the cost-effectiveness of WCD combined with guideline-directed medical therapy (GDMT) compared to GDMT alone, after MI in the English National Health Service (NHS).

View Article and Find Full Text PDF

Purpose: Increasing life expectancy and advances in cancer treatment will lead to more patients needing both radiation therapy (RT) and cardiac implantable electronic devices (CIEDs). CIEDs, including pacemakers and defibrillators, are essential for managing cardiac arrhythmias and heart failure. Telemetric monitoring of CIEDs checks battery status, lead function, settings, and diagnostic data, thereby identifying software deviations or damage.

View Article and Find Full Text PDF

Background: In arrhythmogenic cardiomyopathy (ACM), left ventricle-dominant presentation has poorer outcomes than right-dominant presentation, suggesting that interventricular functional disparity might play a role in patients' prognosis. However, the prognostic impact of ventricular functional discordance in ACM patients remains unknown.

Purpose: To assess whether ventricular functional disparity measured as ventricular discordance index, defined as the ratio of right-ventricular ejection fraction (RVEF) to left-ventricular ejection fraction (LVEF), might reveal prognostic disparities between phenotypes and offer added risk stratification value.

View Article and Find Full Text PDF

Background: Noninferiority of omitting intraoperative defibrillation threshold (DFT) testing has been documented for transvenous implantable cardioverter defibrillators (ICD) whereas data for the subcutaneous-ICD (S-ICD) regarding the need for DFT testing, especially during S-ICD generator replacement, is not available.

Methods: A total of 112 consecutive patients who underwent S-ICD generator replacement and routine testing were included in this retrospective single-center study and analyzed regarding the outcome of intraoperative DFT.

Results: The majority of patients (87.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!