Introduction: The sensing performance of transvenous lead systems may be adversely affected by the delivery of high-energy shocks. This may be due to the proximity of the sensing and energy-delivery electrodes on transvenous leads.
Methods And Results: The time required for detection of ventricular fibrillation and redetection after a failed first shock was compared in 93 patients with five different lead system-pulse generator combinations: Cadence--Endotak 60 series, Ventak P--Endotak 60 series, Jewel--Transvene, Cadence--TVL, and Cadence--Transvene. A total of 418 successful and 204 failed first shocks were delivered during induced ventricular fibrillation. Redetection times (RED) were consistently shorter than detection times (DET) in the Jewel-Transvene (RED minus DET: -1.9 +/- 0.8 sec, P < 0.0001), the Cadence-TVL (-1.6 +/- 1.0 sec, P < 0.0001), and the Cadence-Transvene combinations (-2.0 +/- 0.9 sec, P < 0.0004). Redetection times were not significantly different than detection times in the Cadence-Endotak combination (0.9 +/- 3.1 sec; P = 0.09). Redetection times were significantly longer than detection times in the Ventak-Endotak combination (1.2 +/- 2.3 sec; P = 0.034). Prolonged individual redetection episodes (> 8.2 sec) were observed in the Cadence-Endotak (7 [10%] of 73 episodes) and the Ventak-Endotak (4 [10%] of 39 episodes), but not in the Jewel-Transvene, the Cadence-TVL, and the Cadence-Transvene combinations.
Conclusions: Redetection of ventricular fibrillation may be delayed in some transvenous lead-pulse generator combinations. Successful redetection of ventricular fibrillation following a failed first shock should be demonstrated prior to hospital discharge of patients with implantable defibrillators.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8167.1995.tb00436.x | DOI Listing |
Cardiovasc Revasc Med
January 2025
Department of Cardiovascular disease, Henry Ford, Detroit, MI, USA.
Introduction: Cardiogenic shock (CS) is marked by substantial morbidity and mortality. The two major CS etiologies include heart failure (HF) and acute myocardial infarction (AMI). The utilization trends of mechanical circulatory support (MCS) and their clinical outcomes are not well described.
View Article and Find Full Text PDFAm J Med
January 2025
Professor of Medicine, Department of Cardiology, Tufts Medical Center, (Tufts University School of Medicine), Boston, MA 02111, USA. Electronic address:
Cardiologists and gastroenterologists often encounter the coexistence of symptoms and functional abnormalities, but determining causation is more difficult. In 1962 Smith and Papp first coined the term "linked angina". Their statement was preceded by the experiment whereby increase in bile duct pressure elicited the typical chest pain in patients with ischemic heart disease.
View Article and Find Full Text PDFJ Clin Sleep Med
January 2025
Division of Sleep Medicine, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania.
Study Objectives: Evaluate the performance of the SANSA device to simultaneously assess obstructive sleep apnea (OSA) and cardiac arrhythmias.
Methods: Participants suspected or known to have OSA underwent polysomnography (PSG) while wearing SANSA. SANSA's algorithm was trained using 86 records and tested on 67 to evaluate training bias.
J Interv Card Electrophysiol
January 2025
Rooney Heart Institute, 311 9th St N #201, Naples, FL, 34102, USA.
Introduction: The role of the sympathetic nervous system in the initiation and continuation of ventricular tachyarrhythmias (VTA) is well established. However, whether CSD reduces implantable cardioverter-defibrillator (ICD) shocks and recurrent VTA is still uncertain.
Methods: A comprehensive literature search was performed at Medline and Embase until March 2023.
Future Cardiol
January 2025
Echocardiography research Center, Rajaie cardiovascular medical and research Center, Iran University of Medical Science, Tehran, Iran.
Introduction: Decreased left atrial appendage emptying velocity (LAAV) is a marker for thrombus formation. This study evaluates the association between LAAV and inflammatory indices in non-valvular atrial fibrillation (AF) patients.
Methods: The study population was 1428 patients with AF, 875 of whom enrolled.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!