We studied the response time (RT) and reliability of three neonatal patient-triggered ventilator (PTV) systems: the Draeger Babylog 8000, the Bear Cub enhancement module (CEM), and the Infrasonics Star Sync. In 10 adult rabbits, airway flow and pressure recordings showed the RT of the Star Sync to be shorter than that of the Bear CEM (53 +/- 13 versus 65 +/- 15 ms, p < 0.05), and both were shorter than that of the Babylog (95 +/- 24 ms, p < 0.01) by ANOVA. The RT of the Bear CEM and the Babylog increased significantly at decreased trigger sensitivity settings. All ventilators triggered successfully on assist-control (A/C). However, the Babylog had a higher rate of asynchrony on SIMV (30 +/- 25%) than the Bear CEM (1.1 +/- 0.3%) and the Star Sync (1.2 +/- 0.4%), p < 0.01. In 10 infants with respiratory failure, recordings of airway flow and pressure were made at ventilator inspiratory time (Ti) settings of 0.3, 0.4, and 0.5 s on assist-control and on SIMV at rates of 15, 30, 45, and 60 breaths/min. The Star Sync and Bear CEM triggered successfully on A/C (100%) and had low rates of asynchrony on SIMV (1 to 3%). The Babylog had a lower success rate on A/C (70 +/- 12%) and a higher rate of asynchrony on SIMV (29 +/- 30%) than the other two ventilators; p < 0.01. The lower reliability of the Babylog was due to its variable refractory period (0.2 to 0.5 s, to equal the set Ti).(ABSTRACT TRUNCATED AT 250 WORDS)

Download full-text PDF

Source
http://dx.doi.org/10.1164/ajrccm/148.2.358DOI Listing

Publication Analysis

Top Keywords

star sync
16
bear cem
16
asynchrony simv
12
response time
8
time reliability
8
reliability three
8
three neonatal
8
neonatal patient-triggered
8
airway flow
8
flow pressure
8

Similar Publications

Flatbands appear in many condensed matter systems, including the two-dimensional electron gas in a high magnetic field, correlated materials, and moiré heterostructures. They are characterized by intrinsic geometric properties such as the Berry curvature and Fubini-Study metric. The influence of the band geometry on electron-electron interaction is difficult to understand analytically because the geometry is in general nonuniform in momentum space.

View Article and Find Full Text PDF

We studied the response time (RT) and reliability of three neonatal patient-triggered ventilator (PTV) systems: the Draeger Babylog 8000, the Bear Cub enhancement module (CEM), and the Infrasonics Star Sync. In 10 adult rabbits, airway flow and pressure recordings showed the RT of the Star Sync to be shorter than that of the Bear CEM (53 +/- 13 versus 65 +/- 15 ms, p < 0.05), and both were shorter than that of the Babylog (95 +/- 24 ms, p < 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!