Purpose: To determine the efficacy of a disinfectant wipe method in eliminating adenovirus 8 from both Goldmann tonometer and pneumotonometer tips, and to determine the efficacy of 5-minute disinfectant soaks in removing the virus from Goldmann tonometer tips.
Method: Quantification of adenovirus 8 in the ocular secretions of six infected patients was performed. Clinically relevant inocula of adenovirus 8 were then prepared in a serum suspension which was used to contaminate Goldmann tonometer and pneumotonometer tips. Each contaminated tip was subjected to either a dry wipe or a wipe with one of the following: water, isopropyl alcohol, hydrogen peroxide, and iodophor. The tips were than quantitatively assayed for residual adenovirus 8. Disinfectant soaks were similarly assessed using contaminated Goldmann tonometer tips soaked in water, hydrogen peroxide, iodophor, and sodium hypochlorite.
Results: Adenovirus 8 was recovered from both types of tonometer tips after dry wipes, but no viable virus was recovered from tips wiped with water only or with any of the disinfectants tested. Adenovirus 8 also was recovered from Goldmann tonometer tips that were either rinsed or soaked in water; no virus was recovered from tips soaked for 5 minutes in any of the disinfectants tested.
Conclusion: A disinfectant wipe method using isopropyl alcohol, hydrogen peroxide, or iodophor is effective in removing adenovirus 8 from both Goldmann tonometer and pneumotonometer tips. Five-minute disinfectant soaks also are effective for Goldmann tonometer tips.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0161-6420(93)31388-6 | DOI Listing |
Clin Exp Ophthalmol
January 2025
Eye Clinic, Wellington Regional Hospital, Wellington, New Zealand.
Graefes Arch Clin Exp Ophthalmol
December 2024
Kaiya Eye Clinic, Hamamatsu, Shizuoka, Japan.
Purpose: Little is known about the effects of ambient atmospheric pressure (AP) on intraocular pressure (IOP) under ordinary conditions. This study aimed to investigate the effects of AP on Goldmann applanation tonometer-measured IOP (GAT-IOP) in normal eyes under everyday atmospheric conditions adjusting for effects of possible confounding factors including other climatic factors.
Methods: Data obtained from 2,431 normal healthy eyes of 2,431 subjects (mean age: 56.
J Clin Med
November 2024
2nd Department of Ophthalmology, School of Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki, 54124 Thessaloniki, Greece.
Intraocular pressure (IOP) readings using three different methods (Goldmann applanation tonometry (GAT), Corvis ST, and iCare) were compared in patients who underwent penetrating keratoplasty (PK). : An observational cross-sectional study with prospective recruitment of patients was conducted. IOP measurements were acquired using GAT, iCare, and Corvis (including both uncorrected IOP (CVS-IOP) and biomechanical IOP (bIOP)), and the agreement among methods was analyzed using Bland-Altman plots.
View Article and Find Full Text PDFOphthalmol Glaucoma
November 2024
Department of Ophthalmology, University of Sao Paulo School of Medicine, Sao Paulo, São Paulo 01246-903, Brazil.
Purpose: To investigate the agreement between the Icare HOME tonometer and the Goldmann Applanation Tonometry (GAT) in assessing the Peak Intraocular Pressure during the Water-Drinking Test (WDT).
Design: Prospective cross-sectional study.
Subjects: 79 eyes of 41 open-angle glaucoma patients were included in the study.
Ophthalmol Glaucoma
November 2024
Glaucoma Services, Aravind Eye Hospital, Madurai, India.
Topic: The systematic review and meta-analysis consolidates the extant body of evidence comparing reliability and agreement between Transpalpebral Tonometers (TT) and Goldmann Applanation Tonometer (GAT).
Clinical Relevance: With a global prevalence of 3.54 percent, glaucoma stands as the second leading cause of preventable blindness.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!