Purpose: To determine the efficacy of a disinfectant wipe method in eliminating adenovirus 8 from both Goldmann tonometer and pneumotonometer tips, and to determine the efficacy of 5-minute disinfectant soaks in removing the virus from Goldmann tonometer tips.

Method: Quantification of adenovirus 8 in the ocular secretions of six infected patients was performed. Clinically relevant inocula of adenovirus 8 were then prepared in a serum suspension which was used to contaminate Goldmann tonometer and pneumotonometer tips. Each contaminated tip was subjected to either a dry wipe or a wipe with one of the following: water, isopropyl alcohol, hydrogen peroxide, and iodophor. The tips were than quantitatively assayed for residual adenovirus 8. Disinfectant soaks were similarly assessed using contaminated Goldmann tonometer tips soaked in water, hydrogen peroxide, iodophor, and sodium hypochlorite.

Results: Adenovirus 8 was recovered from both types of tonometer tips after dry wipes, but no viable virus was recovered from tips wiped with water only or with any of the disinfectants tested. Adenovirus 8 also was recovered from Goldmann tonometer tips that were either rinsed or soaked in water; no virus was recovered from tips soaked for 5 minutes in any of the disinfectants tested.

Conclusion: A disinfectant wipe method using isopropyl alcohol, hydrogen peroxide, or iodophor is effective in removing adenovirus 8 from both Goldmann tonometer and pneumotonometer tips. Five-minute disinfectant soaks also are effective for Goldmann tonometer tips.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0161-6420(93)31388-6DOI Listing

Publication Analysis

Top Keywords

goldmann tonometer
28
tonometer tips
20
disinfectant wipe
12
wipe method
12
tonometer pneumotonometer
12
pneumotonometer tips
12
disinfectant soaks
12
hydrogen peroxide
12
peroxide iodophor
12
tips
11

Similar Publications

Purpose: Little is known about the effects of ambient atmospheric pressure (AP) on intraocular pressure (IOP) under ordinary conditions. This study aimed to investigate the effects of AP on Goldmann applanation tonometer-measured IOP (GAT-IOP) in normal eyes under everyday atmospheric conditions adjusting for effects of possible confounding factors including other climatic factors.

Methods: Data obtained from 2,431 normal healthy eyes of 2,431 subjects (mean age: 56.

View Article and Find Full Text PDF

Intraocular pressure (IOP) readings using three different methods (Goldmann applanation tonometry (GAT), Corvis ST, and iCare) were compared in patients who underwent penetrating keratoplasty (PK). : An observational cross-sectional study with prospective recruitment of patients was conducted. IOP measurements were acquired using GAT, iCare, and Corvis (including both uncorrected IOP (CVS-IOP) and biomechanical IOP (bIOP)), and the agreement among methods was analyzed using Bland-Altman plots.

View Article and Find Full Text PDF

Purpose: To investigate the agreement between the Icare HOME tonometer and the Goldmann Applanation Tonometry (GAT) in assessing the Peak Intraocular Pressure during the Water-Drinking Test (WDT).

Design: Prospective cross-sectional study.

Subjects: 79 eyes of 41 open-angle glaucoma patients were included in the study.

View Article and Find Full Text PDF

Topic: The systematic review and meta-analysis consolidates the extant body of evidence comparing reliability and agreement between Transpalpebral Tonometers (TT) and Goldmann Applanation Tonometer (GAT).

Clinical Relevance: With a global prevalence of 3.54 percent, glaucoma stands as the second leading cause of preventable blindness.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!