Standard silver-based films are usually too sensitive to be used as direct indicators of dose in dynamic radiosurgery because of optical saturation. This paper describes the use of a new radiochromic film to measure 6-MV radiosurgery doses and dose distributions in a head phantom. Dose calibration of the radiochromic film was performed in the range of 2.3-50.2 Gy using light of 632- and 530-nm wavelengths. Radiosurgery dose distributions were measured using the radiochromic film in a head phantom undergoing the same treatment as a patient, and were compared with the planned distributions. For an example case (nominal 2.0-cm-diam cone), film measurement verified the calculated dose distribution in one plane. The simple measurement technique described led to experimental uncertainties of +/- 0.1 cm for the 90% and 50% isodose lines, +/- 0.3 cm for the 20% line, and +/- 0.5 cm for the 10% line. Isocenter dose was measured with an uncertainty of +/- 3%. Refinements to the technique should allow more precise measurements. It is concluded that the radiochromic film, with some limitations, is a convenient and useful tool for dynamic radiosurgery quality assurance.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.597385DOI Listing

Publication Analysis

Top Keywords

radiochromic film
20
dynamic radiosurgery
8
dose distributions
8
head phantom
8
dose
6
radiochromic
5
radiosurgery
5
film
5
film treatment
4
treatment verification
4

Similar Publications

Strontium-90 plesiotherapy delivers high doses of radiation to superficial lesions (<3 mm depth) with excellent sparing of deeper tissues. The sealed-source applicator tip is circular and 8-10 mm in diameter. Larger treatment fields are treated with multiple overlapping fields.

View Article and Find Full Text PDF

Purpose: The spot size of scanned particle beams is of crucial importance for the correct dose delivery and, therefore, plays a significant role in the quality assurance (QA) of pencil beam scanning ion beam therapy.

Materials And Methods: This study compares 5 detector types-radiochromic film, ionization chamber (IC) array, flat panel detector, multiwire chamber, and IC-for measuring the spot size of proton and carbon ion beams.

Results: Variations of up to 30% were found between detectors, underscoring the impact of detector choice on QA outcomes.

View Article and Find Full Text PDF

Purpose: In radiotherapy treatment planning systems, modelling of superficial dose may be aided by a body contour that is, by default, placed at the outermost air-tissue interface. Here we investigate the accuracy of superficial dose calculated using either the default body contour (DBC) or an extended body contour (EBC) compared to radiochromic film measurements made on a slab phantom and an anthropomorphic phantom.

Methods: Depth dose curves in the superficial region of the slab phantom were measured using stacked radiochromic films and irradiated using static beams delivered from varying incident angles.

View Article and Find Full Text PDF

This study elucidated the radiation response characteristics of a Gafchromic radiochromic film subjected to low photon doses of ≤50 mSv, which corresponds to the annual whole body effective dose limit for radiation workers in Canada. Radiochromic films are investigated for possible use as a complementary tool for the Canadian Armed Forces that can be worn in addition to their existing personal dosimetry to quickly assess personal radiation dose received from radiological hazards without reliance on electronics. The films were exposed to varying photon energies emanating from x-ray generators and radioisotopes, specifically cesium-137, cobalt-60, and americium-241.

View Article and Find Full Text PDF

This study analyzed the spectral response of EBT3, EBT4, and EBT-XD radiochromic films using absorption spectroscopy. The primary focus was on characterizing the evolution of spectral signatures across a range of absorbed doses, thereby elucidating the unique dose-dependent response profiles of each film type. Ten samples of each film type were subjected to open field irradiation within their designated dose ranges (1-20 Gy for EBT3 and EBT4, 1-50 Gy for EBT-XD).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!